State v. Consolo ( 2013 )


Menu:
  • [Cite as State v. Consolo, 2013-Ohio-2611.]
    IN THE COURT OF APPEALS
    ELEVENTH APPELLATE DISTRICT
    PORTAGE COUNTY, OHIO
    STATE OF OHIO,                                  :         OPINION
    Plaintiff-Appellant,           :
    CASE NO. 2012-P-0106
    - vs -                                  :
    WILLIAM F. CONSOLO,                             :
    Defendant-Appellee.            :
    Criminal Appeal from the Portage County Municipal Court, Ravenna Division, Case No.
    R2011 TRC 16608.
    Judgment: Reversed and remanded.
    Victor V. Vigluicci, Portage County Prosecutor, and Pamela J. Holder, Assistant
    Prosecutor, 241 South Chestnut Street, Ravenna, OH 44266 (For Plaintiff-Appellant).
    Edwin R. O’Day, 167 Royal Oak Drive, Aurora, OH 44202, and Joseph C. Patituce,
    Patituce & Associates, LLC, 26777 Lorain Road, Suite 708, North Olmsted, OH 44070
    (For Defendant-Appellee).
    DIANE V. GRENDELL, J.
    {¶1}     Plaintiff-appellant, the State of Ohio, appeals the August 30, 2012 Journal
    Entry of the Portage County Municipal Court, Ravenna Division, suppressing the
    defendant-appellee, William F. Consolo, Jr.’s breath test results. The issues before this
    court are whether the State bears the burden of producing evidence of the general,
    scientific reliability of the Intoxilyzer 8000 when challenged in a motion to suppress;
    whether the Ohio Administrative Code requires a dry gas control between subject tests;
    and whether the alleged alteration of Consolo’s breath test results by the Ohio
    Department of Health on its website is reason to suppress the results. For the following
    reasons, we reverse the decision of the court below.
    {¶2}   On December 19, 2011, the Streetsboro Police Department issued
    Consolo a traffic ticket, charging him with OVI, a misdemeanor of the first degree in
    violation of R.C. 4511.19(A)(1)(a) (driving under the influence of alcohol) and (d) (driving
    with a prohibited breath alcohol concentration), and with violating the Approaching a
    Stopped Emergency Vehicle statute, a minor misdemeanor in violation of R.C.
    4511.213.
    {¶3}   On December 23, 2011, Consolo entered a plea of not guilty.
    {¶4}   On February 21, 2012, Consolo filed a Motion to Suppress, seeking, inter
    alia, to suppress the breath test results of the Intoxilyzer 8000.         As grounds for
    suppression, Consolo argued that the breath test was not conducted in accordance with
    the regulations of the Ohio Department of Health as set forth in Chapter 3701-53-02 of
    the Ohio Administrative Code, and the Intoxilyzer 8000 was determined by the
    municipal court to be unreliable in State v. Johnson, Portage County Municipal Court,
    Ravenna Division, Case No. R 2011 TRC 4090.
    {¶5}   On May 11, 2012, Consolo filed a Supplement to Motion to Suppress,
    arguing, as an additional reason to suppress the breath test results, that the samples
    were not taken in accordance with Ohio Adm.Code 3701-53-04(B), which requires “a
    dry gas control test before and after every subject test.”
    {¶6}   On August 29, 2012, a suppression hearing was held. Mary Martin, the
    Program Administrator for Alcohol and Drug Testing at the Ohio Department of Health,
    testified on behalf of the State. Martin testified that she was trained in the operation of
    2
    the Intoxilyzer 8000 by the manufacturer, and is responsible for developing regulations
    for its use.
    {¶7}      Martin explained that a “dry gas control” is a “known mixture of ethanol * *
    * that has a known target value[.] * * * If the instrument can read the dry gas control
    within an appropriate range * * * that shows that the instrument is in proper working
    order.”1
    {¶8}      With respect to Ohio Adm.Code 3701-53-04(B), Martin testified that “there
    is one dry gas control at the beginning of the subject test and one at the end of the
    subject test.”
    {¶9}      Consolo’s Subject Test Report was introduced into evidence. The Report
    contained the following Test Information:
    Test                        BrAC (g/210L)                  Time
    Air Blank                   0.000                          02:57
    Diagnostic                  VAC/OK                         02:57
    Air Blank                   0.000                          02:58
    Dry Gas Control             0.100                          02:58
    Air Blank                   0.000                          02:58
    Subject Test 1              0.162                          03:00
    Air Blank                   0.000                          03:00
    Air Blank                   0.000                          03:03
    Subject Test 2              0.146                          03:03
    Air Blank                   0.000                          03:04
    Dry Gas Control             0.100                          03:04
    Air Blank                   0.000                          03:05
    {¶10} Martin testified that the term “subject test,” as used in Ohio Adm.Code
    3701-53-04(B), is undefined. According to Martin, a subject test “is basically the whole
    document,” i.e., the Subject Test Report, and not Subject Tests 1 and 2 contained within
    1. According to the Bureau of Alcohol and Drug Testing’s training manual for the Intoxilyzer 8000,
    introduced as evidence at the suppression hearing: “The instrument automatically performs a DRY GAS
    CONTROL; The DRY GAS CONTROL checks the accuracy of the instrument before the subject test.
    3
    the Report. Thus, only two dry gas control tests are required and it is not necessary to
    perform dry gas controls between Subject Test 1 and Subject Test 2.                        Martin’s
    understanding of what constitutes a “subject test” for the purposes of Ohio Adm.Code
    3701-53-04(B) is consistent with the Bureau of Alcohol and Drug Testing’s training
    manual for the Intoxilyzer 8000.
    {¶11} Martin further testified that Subject Tests 1 and 2 are readings of actual
    breath samples taken from the test subject, i.e., Consolo. She testified that “you have
    to have two subject samples to have a valid subject test,” and that “the lower of the two
    [is] the chargeable per se level.”
    {¶12} Finally, Martin testified that, because of confusion as to whether the
    subject test contained in Ohio Adm.Code 3701-53-04(B) referred to the Subject Test
    Report or Subject Test 1 and 2, she changed the wording in the Report. The terms
    Subject Test 1 and Subject Test 2 were replaced with Subject Sample 1 and Subject
    Sample 2. Although Consolo’s Subject Test Report, generated on December 19, 2011,
    contains the terms Subject Test 1 and 2, the Subject Test Report posted on the Ohio
    Department of Health’s website uses the terms Subject Sample 1 and 2.2 The BrAC
    readings are the same on both Test Reports.
    {¶13} On August 30, 2012, the municipal court issued a Journal Entry, granting
    Consolo’s Motion to Suppress with respect to the breath test results produced by the
    Intoxilyzer 8000.     Relying on the Johnson decision, the court held that, “when a
    defendant raises a general, scientific attack on the reliability of the Intoxilyzer 8000, it is
    the State’s burden to produce certain evidence to convince the Court that the results
    The DRY GAS CONTROL target value is 0.100; If the DRY GAS CONTROL results are outside the range
    of +/- 0.005; (Lower than 0.095 or higher than 0.105) The test will ABORT.”
    2. http://publicapps.odh.ohio.gov/BreathInstrument/Testreport.aspx?id=84300 (accessed June 12, 2013).
    4
    from the Intoxilyzer 8000 are scientifically reliable, credible and competent evidence.”
    Since the State did not produce such evidence, the breath test results would not be
    admitted at trial.
    {¶14} The municipal court further held that Ohio Adm.Code 3701-53-04(B) “is
    clear and is not ambiguous in any way,” and that it required a dry gas control to be
    performed before and after every subject test. Since there was no dry gas control
    performed between Subject Test 1 and Subject Test 2, the breath test results would not
    be admitted at trial.
    {¶15} Finally, the municipal court found that “the Ohio Department of Health has
    unilaterally changed the records of Defendant’s breath test.”
    {¶16} For the foregoing reasons, the municipal court dismissed the violation of
    R.C. 4511.19(A)(1)(d) (driving with a prohibited breath alcohol concentration), and
    ordered the remaining two counts to be set for trial.
    {¶17} On September 5, 2012, the State filed its Notice of Appeal and a Motion to
    Stay the Execution of Judgment.
    {¶18} On appeal, the State raises the following assignments of error:
    {¶19} “[1.] [The] Portage County Municipal Court erred in permitting a general
    attack on the scientific reliability of the Intoxilyzer 8000 contrary to Ohio statutes and
    well-established case law.”
    {¶20} “[2.] The Portage County Municipal Court erred in finding the testing
    protocol for the Intoxilyzer 8000 did not comply with Ohio Adm.Code 3701-53[-]04(B),
    rendering the test results inadmissible at trial.”
    5
    {¶21} “[3.] The Portage County Municipal Court erred in excluding the results of
    the Intoxilyzer 8000 test based on the court’s finding that the records of Consolo’s
    breath test results had been changed by the Ohio Department of Health.”
    {¶22} The appropriate standard of review where the lower court’s judgment is
    challenged on a purported misconstruction of the law is de novo. State v. Morris, 
    132 Ohio St. 3d 337
    , 2012-Ohio-2407, 
    972 N.E.2d 528
    , ¶ 16.            “In determining a pure
    question of law, an appellate court may properly substitute its judgment for that of the
    trial court.” (Citation omitted.) 
    Id. {¶23} Where,
    however, the issue requires the consideration of factual evidence,
    a reviewing court must defer to the trial court’s factual findings, if supported by
    competent, credible evidence. State v. Mayl, 
    106 Ohio St. 3d 207
    , 2005-Ohio-4629, 
    833 N.E.2d 1216
    , ¶ 41. The reviewing court will then independently determine whether the
    facts satisfy the applicable legal standard. Id.; Kirtland Hills v. Kunka, 11th Dist. No.
    2012-L-095, 2013-Ohio-738, ¶ 16 (“[o]nce the appellate court accepts the trial court’s
    factual determinations, the appellate court conducts a de novo review of the trial court’s
    application of the law to these facts”) (citation omitted).
    {¶24} In its first assignment of error, the State argues that, pursuant to Sections
    3701.143 and 4511.19(D)(1)(b) of the Ohio Revised Code, and Ohio Administrative
    Code 3701-53-02(A)(3), a trial court is required to accept the Intoxilyzer 8000 as an
    appropriate device for chemically analyzing a person’s breath to ascertain the amount of
    alcohol in the breath.
    {¶25} The Johnson case, relied upon by the municipal court, has been reversed.
    State v. Johnson, 11th Dist. No. 2012-P-0008, 2013-Ohio-440, ¶ 32 (“the legislature has
    allowed the director of the department of health to determine that the Intoxilyzer 8000 is
    6
    generally reliable”). This court has held, in many other decisions, that the State is not
    required to introduce evidence of the Intoxilyzer 8000’s scientific reliability for the breath
    test results to be admissible. State v. Miller, 11th Dist. No. 2012-P-0032, 2012-Ohio-
    5585, ¶ 26, quoting State v. Vega, 
    12 Ohio St. 3d 185
    , 190, 
    465 N.E.2d 1303
    (1984) (“an
    accused may not make a general attack upon the reliability and validity of the breath
    testing instrument”).
    {¶26} The first assignment of error is with merit.
    {¶27} In the second assignment of error, the State argues that the municipal
    court erred in its interpretation of Ohio Adm.Code 3701-53-04(B). We agree.
    {¶28} “In the absence of any definition of the intended meaning of words or
    terms used in a legislative enactment they will, in the interpretation of the act, be given
    their common, ordinary and accepted meaning in the connection in which they are
    used.”     Baker v. Powhatan Mining Co., 
    146 Ohio St. 600
    , 
    67 N.E.2d 714
    (1946),
    paragraph three of the syllabus; R.C. 1.42 (“[w]ords and phrases shall be read in
    context and construed according to the rules of grammar and common usage”). “This
    maxim applies equally to administrative regulations.” State ex rel. Brilliant Elec. Sign
    Co. v. Indus. Comm. of Ohio, 
    57 Ohio St. 2d 51
    , 54, 
    386 N.E.2d 1107
    (1979).
    {¶29} In the interpretation of administrative regulations, “considerable deference
    should be accorded to an agency’s interpretation of rules the agency is required to
    administer.” State ex rel. Celebrezze v. Natl. Lime & Stone Co., 
    68 Ohio St. 3d 377
    ,
    382, 
    627 N.E.2d 538
    (1994); Swallow v. Indus. Comm. of Ohio, 
    36 Ohio St. 3d 55
    , 57,
    
    521 N.E.2d 778
    (1988) (“courts * * * must give due deference to an administrative
    interpretation formulated by an agency that has accumulated substantial expertise, and
    to which the General Assembly has delegated the responsibility of implementing the
    7
    legislative command”); compare Chevron U.S.A., Inc. v. Natural Resources Defense
    Council, Inc., 
    467 U.S. 837
    , 843, 
    107 S. Ct. 2778
    , 
    81 L. Ed. 2d 694
    (1984) (“if the statute
    is silent or ambiguous with respect to the specific issue, the question for the court is
    whether the agency’s answer is based on a permissible construction of the statute”).
    {¶30} “[T]he director of health shall determine, or cause to be determined,
    techniques or methods for chemically analyzing a person’s * * * breath * * * in order to
    ascertain the amount of alcohol * * * in the person’s * * * breath.” R.C. 3701.143.
    Pursuant to this statute, the Director of Health has determined that the Intoxilyzer 8000
    “shall automatically perform a dry gas control test before and after every subject test
    and instrument certification using a dry gas standard traceable to the national institute of
    standards and technology (NIST).” Ohio Adm.Code 3701-53-04(B).
    {¶31} The issue herein is whether the term “subject test” refers to the test of an
    individual yielding the final result, or whether it refers to the two breath samples
    necessary to yield the final result. Any confusion on this issue does not derive from the
    words of the regulation, but from the fact that the Subject Test Form designated the two
    breath samples as “Subject Test 1” and “Subject Test 2.” Such confusion is readily
    dispelled in light of the Department of Health’s interpretation of its own regulation.
    According to Martin’s uncontradicted testimony, a dry gas control is to be performed
    twice, once at the beginning of the testing procedure and again at the end. This is the
    way the manufacturer trained Martin in the Intoxilyzer 8000’s operation, and, in turn, this
    is the way the Department of Health has trained law enforcement personnel. We further
    note that the dry gas control is performed automatically as part of the testing sequence,
    i.e., the operator does not initiate the dry gas control, rather, the Intoxilyzer is
    8
    programmed to operate so that only two dry gas controls are performed during each
    subject test.
    {¶32} Other appellate courts that have considered this issue have reached the
    same conclusion. In State v. Kormos, 12th Dist. No. CA2011-08-059, 2012-Ohio-3128,
    the court of appeals reasoned:
    {¶33} [I]t is quite clear that a “subject test” is synonymous with running
    the Intoxilyzer 8000 on a single “subject,” i.e., “individual” under law
    enforcement’s control.     Because the language set forth by the
    Health Department is clear and definite, we must apply it as written
    and hold that there is only one “subject” being tested during any
    one breathalyzer test. See Kneisley v. Lattimer-Stevens Co., 
    40 Ohio St. 3d 354
    , 357, 
    533 N.E.2d 743
    (1988). It follows that the
    items listed in the “Subject Test Report” form, including “Subject
    Test 1” and “Subject Test 2,” are simply recurring components of
    the same breathalyzer test, where one subject blows twice before
    reaching the end result.
    
    Id. at ¶
    16. Accord Cincinnati v. Nicholson, 1st Dist. No. C-120332, 2013-Ohio-708, ¶
    11.
    {¶34} The second assignment of error is with merit.
    {¶35} In the third assignment of error, the State argues that the municipal court
    erred in its determination that a change in the data displayed on the Ohio Department of
    Health’s website constituted a change in Consolo’s records, warranting their
    suppression. Again, we agree.
    9
    {¶36} The issue of what information the Ohio Department of Health chooses to
    publish on its website, and how that information is displayed, is irrelevant to the issue of
    whether the bodily substance was “analyzed in accordance with methods approved by
    the director of health” and “by an individual possessing a valid permit.”               R.C.
    4511.19(D)(1)(b). Consolo’s original Subject Test Report, generated on December 19,
    2011, remains unaltered.       This is the only relevant Test Report for establishing
    Consolo’s BrAC.
    {¶37} The third assignment of error is with merit.
    {¶38} For the forgoing reasons, the judgment of the Portage County Municipal
    Court, Ravenna Division, suppressing Consolo’s breath test results and dismissing the
    OVI charge for violating R.C. 4511.19(A)(1)(d) (driving with a prohibited breath alcohol
    concentration), is reversed, and this cause is remanded for further proceedings
    consistent with this opinion. Costs to be taxed against appellee.
    CYNTHIA WESTCOTT RICE, J., concurs,
    THOMAS R. WRIGHT, J., dissents with a Dissenting Opinion.
    _____________________
    THOMAS R. WRIGHT, J., dissents with a Dissenting Opinion.
    {¶39} R.C. 4511.19(D)(1)(b) does not mandate admissibility of breath test
    results derived from the Intoxilyzer 8000.        Rather, that statute which, by its plain
    language controls the issue in this case, vests the trial court with discretion regarding
    admissibility despite approval from the director. I, therefore, respectfully dissent.
    10
    {¶40} R.C. 3701.143 empowers the director to approve breath testing devices,
    and R.C. 4511.19(D)(1)(b) grants trial courts the discretion to admit the results from
    approved devices without further proof of reliability when circumstances warrant.
    Although some claim the contrary, nobody is correct all the time. In recognizing human
    fallibility, the legislature had the wisdom to vest within the trial court the discretion per
    R.C.4511.19(D)(1)(b) to conduct further inquiry when there is an issue as to the
    reliability of an approved breath testing device before admitting the results.
    {¶41} R.C. 4511.19(D)(1)(b) states that “[i]n any criminal prosecution or juvenile
    court proceeding for a violation of division (A) or (B) of this section or for an equivalent
    offense that is vehicle-related, the court may admit evidence on the concentration of
    alcohol, drugs of abuse, controlled substances, metabolites of a controlled substance,
    or a combination of them in the defendant’s whole blood, blood serum or plasma,
    breath, urine, or other bodily substance at the time of the alleged violation as shown by
    chemical analysis of the substance withdrawn within three hours of the time of the
    alleged violation[,]” and “[t]he bodily substance withdrawn under division (D)(1)(b) of this
    section shall be analyzed in accordance with methods approved by the director of
    health by an individual possessing a valid permit issued by the director pursuant to
    section 3701.143 of the Revised Code.” (Emphasis added.)
    {¶42} The statute does not use the word “shall,” which would mandate
    admission regardless of the circumstances. Rather, the statute uses the word “may.”
    For purposes of statutory construction, “use of the word ‘may’ is generally construed to
    make the provision in which it is contained optional, permissive, or discretionary * * *.”
    Dorrian v. Scioto Conservancy Dist., 
    27 Ohio St. 2d 102
    , 107 (1971); State v. Suchevits,
    
    138 Ohio App. 3d 99
    , 102 (11th Dist. 1999).
    11
    {¶43} In this case, the trial court exercised its discretion not to admit the breath
    test absent proof from the state that the Intoxilyzer 8000 is generally reliable, a decision
    consistent with the discretion it possesses under R.C.4511.19(D)(1)(b). As reliability
    presents a threshold admissibility issue, reliability, as opposed to the weight to be
    afforded any admitted evidence, is one for the trial court. Knott v Revolution Software
    Inc. 
    181 Ohio App. 3d 519
    , 2009-Ohio-1191, ¶45 (5th Dist.); State v. Riley, 6th Dist. No.
    WD-03-076, 2007-Ohio-879, ¶27 (expert testimony must be deemed reliable before it is
    deemed admissible.); Saad v. Shimano American Corp., 
    2000 U.S. Dist. LEXIS 10974
    ,
    *7 (N.D. Ill. 2000)(The Supreme Court has made it clear that the courts must allow into
    evidence only expert testimony that meets certain threshold standards of reliability and
    usefulness).
    {¶44} Moreover, the determination of evidential reliability necessarily implicates
    the defendant’s substantive due process rights.
    {¶45} “Substantive due process, [although an] ephemeral concept, protects
    specific fundamental rights of individual freedom and liberty from deprivation at the
    hands of arbitrary and capricious government action. The fundamental rights protected
    by substantive due process arise from the Constitution itself and have been defined as
    those rights which are ‘implicit in the concept of ordered liberty.’ (* * *) While this is
    admittedly a somewhat vague definition, it is generally held that an interest in liberty or
    property must be impaired before the protections of substantive due process become
    available.” State v. Small, 162 Ohio App.3d. 375, 2005-Ohio-3813, ¶11 (10th Dist.),
    quoting Gutzwiller v. Fenik, 
    860 F. 2d
    . 1317, 1328 (6th Cir. 1989).
    {¶46} However vague the conceptual parameters of one’s substantive due
    process guarantees may be, the following principle is clear; “[substantive] * * * due
    12
    process is violated by the introduction of seemingly conclusive, but actually unreliable
    evidence.” Barefoot v. Estelle, 
    463 U.S. 880
    , 931, fn. 10 (1983).
    {¶47} The trial court was aware that other courts had deemed the Intoxilyzer
    8000 unreliable even though it was approved. Against this backdrop, the court ordered
    the state to establish the general reliability of the Intoxilyzer 8000 before admitting the
    results. Given the constitutional gravity of admitting unreliable results, however, and its
    statutory authority to act as gatekeeper regarding breath test results, the lower court’s
    decision to require the state to produce evidence of the machines reliability was an
    eminently reasonable and sound legal decision. “[A]n abuse of discretion is the trial
    court’s ‘failure to exercise sound, reasonable, and legal decision-making.’” State v.
    Beechler, 2d Dist. No. 09-CA-54, 2010-Ohio-1900, ¶62, quoting Black’s Law Dictionary
    (8 Ed.Rev.2004) 11.
    {¶48} Rather than present evidence of the general reliability of the Intoxilyzer
    8000, the state took the position that the trial court could not require it to do so pursuant
    to Vega and its progeny. Vega, 
    12 Ohio St. 3d 185
    (1984). I do not read Vega as
    holding that under no circumstances can a trial court exercise its discretion to require
    evidence of general reliability of an approved breath testing device as a condition to
    admissibility.
    {¶49} In Vega, the court held “* * * an accused is not denied his constitutional
    right to present a defense nor is the state relieved of its burden of proving guilt beyond a
    reasonable doubt where a trial judge does not permit expert testimony to attack the
    reliability of intoxilyzers in general.” (Emphasis added.) 
    Id. at 186.
    {¶50} Threshold admissibility was not at issue in Vega. That is, the defendant
    made no challenge to the trial court’s admission of his breath test result. Instead, after
    13
    the state presented its case and rested, the defendant attempted to present a
    “reliability” defense by attacking intoxilyzers in general. See also State v. Vega, 5th
    Dist. No. CA-1766, 1993 Ohio App LEXIS 14350, *16 (Nov.22, 1983)(Hoffman, J.,
    dissenting). Unlike Vega, 
    12 Ohio St. 3d 185
    , threshold admissibility is the issue in the
    case before us.      Moreover, unlike Vega, our case is not about the reliability of
    intoxilyzers in general. Our case is limited to whether the Intoxilyzer 8000 is reliable. In
    short, the circumstances at issue in Vega were fundamentally distinguishable from
    those in our case.
    {¶51} Additionally, the rule in Vega does not contemplate a situation where, as
    here, an approved device’s general reliability has been assessed by other courts for
    both use in and out of this state and the device’s reliability has been found suspect.
    See State v. Johnson, Portage County Municipal Court, January 6, 2012.                Vega
    expressly states that its holding does not involve a situation where the defense asserts
    that there was an abuse of discretion by the director in approving the breath testing
    device at issue. Vega at 187, fn. 2.     Obviously, in our case, if the Intoxilyzer 8000 is
    unreliable, approval would amount to an abuse of discretion and admission of the test
    results a violation of substantive due process.
    {¶52} Breath tests are “‘* * * generally recognized as being reasonably reliable
    on the issue of intoxication when conducted with proper equipment and by competent
    operators.’” (Emphasis added.) Vega at 186, quoting Westerville v. Cunningham, 
    15 Ohio St. 2d 121
    , 128(1968). Thus, the central issue as presented in the case before us,
    does the Intoxilyzer 8000 qualify as “proper equipment”? The answer is “yes” if it is
    generally reliable and “no” if it is not. This is a query, however, that, under Ohio law, a
    trial court is entitled to resolve pursuant to R.C. 4511.19(D)(1)(b).
    14
    {¶53} In this case, the trial court exercised its discretion to safeguard the
    defendant’s right to substantive due process by merely requiring the state to show the
    Intoxilyzer 8000 is generally reliable. Under the circumstances, this decision was sound
    and reasonable.       This is particularly true in light of the fact that a trial court is vested
    with broad discretion in the admission or exclusion of evidence and in recognition that it
    has inherent power to exclude or strike evidence on its own motion. Caroll v Caroll, 7th
    Dist. No. 89-C-1, 1990 Ohio App. LEXIS 1339, *8 (April 5, 1990); Neil v. Hamilton
    County, 
    87 Ohio App. 3d 670
    ; Oakbrook Realty Corp. v. Blout, 
    48 Ohio App. 3d 69
    , 70
    (10th Dist. 1988).
    {¶54} Given the foregoing point, there is no reason to remand this case to the
    trial court based upon perceived inadequacies in the motion to suppress. The trial court
    made it abundantly clear that it would not admit the test results absent proof of reliability
    of the Intoxilyzer 8000. Requiring the proponent to establish the reliability of scientific
    evidence is something that a trial court may require as previously discussed. The state
    was well aware of what the trial court required when it ordered the state to produce
    evidence of the Intoxilyzer 8000’s reliability, independent and irrespective of the
    contents of the motion to suppress. Accordingly, there is no procedural due process
    violation of the state’s right to notice and an opportunity to be heard. The trial court’s
    order was unambiguous and an exercise of the sound discretion as the gatekeeper of
    breath test result admissibility.
    {¶55} When an appellate court [**14] is reviewing a pure issue of law, “the mere
    fact that the reviewing court would decide the issue differently is enough to find error (of
    course, not all errors are reversible. Some are harmless; others are not preserved for
    appellate review). By contrast, where the issue on review has been confined to the
    15
    discretion of the trial court, the mere fact that the reviewing court would have reached a
    different result is not enough, without more, to find error.” Sertz v. Sertz, 11th Dist. No.
    2011-L-063, quoting Beechler, 2010-Ohio-1900 at ¶67.
    {¶56} This appeal is centered around a discretionary decision made by the trial
    court. As I find the court’s decision not only reasonable, but constitutionally astute, I
    would affirm the trial court’s exclusion of the breath test in light of the state’s refusal to
    present evidence on the issue.