  •          IN THE COURT OF CRIMINAL APPEALS OF TENNESSEE
    AT NASHVILLE
    July 15, 2015 Session
    STATE OF TENNESSEE v. JON MICHAEL JOHNSON
    Appeal from the Criminal Court for Davidson County
    No. 2012-B-1415    Amanda McClendon, Judge1
    No. M2014-01834-CCA-R3-CD – Filed October 21, 2015
    Pursuant to Rule 37(b) of the Tennessee Rules of Criminal Procedure, the defendant, Jon
    Michael Johnson, who pleaded guilty to one count of driving under the influence
    (“DUI”), appeals two related certified questions of law relative to the validity of the
    instrument used to measure his blood alcohol level following his arrest. Because neither
    of the certified questions presented is dispositive of the defendant’s case, the appeal is
    dismissed.
    Tenn. R. App. P. 3; Appeal Dismissed
    JAMES CURWOOD WITT, JR., J., delivered the opinion of the court, in which NORMA
    MCGEE OGLE and ALAN E. GLENN, JJ., joined.
    Kevin McGee (on appeal) and James Brian Lewis (on appeal and at trial), Nashville,
    Tennessee, for the appellant, Jon Michael Johnson.
    Herbert H. Slatery III, Attorney General and Reporter; Clark B. Thornton, Assistant
    Attorney General; Victor S. Johnson III, District Attorney General; and Rebecca
    Valiquette, Assistant District Attorney General, for the appellee, State of Tennessee.
    OPINION
    The Davidson County grand jury charged the defendant with one count of
    DUI and one count of driving with a blood alcohol level of .08 percent or greater (“DUI
    per se”). On August 9, 2013, the defendant filed a motion to exclude the results of the
    breath alcohol test conducted at the time of his arrest, arguing that “[t]he Tennessee
    Bureau of Investigation breath test calibration system is not valid science. Therefore, it is
    1
    Judge Mark Fishburn issued the ruling denying the defendant’s motion to suppress the results of
    the blood alcohol testing in this case, which ruling occasioned the certified questions in this case.
    Following Judge Fishburn’s ruling, this misdemeanor DUI case was transferred to Judge McClendon.
    unreliable evidence.” The issue raised in the defendant’s motion was identical to that
    presented in similar motions filed by 12 other defendants. The trial court consolidated
    the motions for the purpose of holding a single pretrial hearing on the issue.
    The trial court conducted a hearing on the consolidated motions on August
    29, 2013. As an initial matter, we note that the definition of several terms is crucial to the
    understanding of the issue presented and the testimony presented at the hearing.
    Accordingly, we provide the definition of those terms as it was provided by Tennessee
    Bureau of Investigation (“TBI”) Special Agent and Forensic Scientist Samera Zavaro,
    supervisor of the “breath alcohol section” of the TBI Crime Laboratory in Nashville, at
    the hearing on the consolidated motions:
    1. “Calibration is when you put a known standard in the
    instrument and tell it what that standard is, so it’s sort of like
    retuning the instrument.”
    2. “Certification is done by running tests through the
    instrument and you know the value but you don’t tell the
    instrument the value and it has to give an answer within [the
    TBI] tolerance range.”
    3. “An accuracy check is by and large the same as a
    certification test[]; a known standard is run through the
    instrument, and if it fails the accuracy check . . . then it will
    not let [the] subject test be performed.” An accuracy check is
    an “internal standard that runs two tests every Sunday at
    11:00, if it’s on. If it’s not on Sunday at 11:00, the next time
    it comes on and it comes to temperature, it will run those two
    tests, that accuracy check.”
    We also note that the breath test machine at issue in this case is the Intoximeter EC/IR II.
    Also, an understanding of the case is enhanced by reviewing the ruling in
    State v. Sensing, the case at the heart of the defendant’s challenge. In Sensing, our
    supreme court, noting that “the results of [breath alcohol] testing where properly
    performed are generally accepted,” relaxed “the rigorous prerequisites formerly required
    to authenticate the reliability of the scientific equipment and procedure when they were
    first employed” and held that it was “no longer necessary for a certified operator of an
    evidentiary breath testing instrument to know the scientific technology involved in the
    function of the machine.” State v. Sensing, 
    843 S.W.2d 412
    , 416 (Tenn. 1992). In place
    -2-
    of the former requirements, the supreme court created a new list of prerequisites for the
    admission of breath alcohol test results at trial:
    We hold that the testing officer must be able to testify (1) that
    the tests were performed in accordance with the standards and
    operating procedure promulgated by the forensic services
    division of the Tennessee Bureau of Investigation, (2) that he
    was properly certified in accordance with those standards, (3)
    that the evidentiary breath testing instrument used was
    certified by the forensic services division, was tested
    regularly for accuracy and was working properly when the
    breath test was performed, (4) that the motorist was observed
    for the requisite 20 minutes prior to the test, and during this
    period, he did not have foreign matter in his mouth, did not
    consume any alcoholic beverage, smoke, or regurgitate, (5)
    evidence that he followed the prescribed operational
    procedure, (6) identify the printout record offered in evidence
    as the result of the test given to the person tested.
    
    Id. at 416.
    The court emphasized that “[t]he breath test result merely creates a rebuttable
    presumption of intoxication” and that, once the State has satisfied the enumerated
    prerequisites, “[t]he defense is then free to rebut the State’s evidence by calling witnesses
    to challenge the accuracy of the particular machine, the qualifications of the operator, and
    the degree to which established testing procedures were followed.” 
    Id. Agent Zavaro
    testified that new breath test machines were initially
    calibrated and certified by the TBI “at a .02, a .05, a .08, a .10[,] and a .20 to show that
    the instrument is reading in a linear fashion.” The 90-day certification is conducted in
    person by a TBI agent using a blank sample followed by a single, .082 percent dry-gas
    standard that is purchased from Intoximeters Incorporated (“Intox”). The weekly
    accuracy checks also use a single .082 dry-gas standard. The machines are not tested for
    linearity after their initial calibration and certification. Agent Zavaro explained that the
    EC/IR II used “an electric chemical cell or a fuel cell” to determine the breath alcohol
    content and that, according to her training, because “the fuel cells are linear, . . . [the
    machine] only requires a one-point calibration, and it’s also common knowledge that
    electrode chemical cells or fuel cells have a linear response.” Agent Zavaro testified that
    Intox told the TBI that the EC/IR II’s fuel cell technology was linear but admitted that
    she did not know whether that information had been subjected to peer review, observing
    only that she knew that information concerning fuel cell linearity was “on Wikipedia.”
    -3-
    Agent Zavaro testified that the TBI did not have a written certification
    protocol in place at the time of any of the cases at issue, including the defendant’s. She
    said that the agency has since drafted a written protocol that is identical to the unwritten
    certification protocol that agents were following previously.
    Macquorn Forrester, Jr., Intox Chief Executive Officer, testified that he had
    assisted in the design of the breath testing machines produced by his family’s company,2
    had written the software for the machines, and had written the training manuals. He said
    that he holds two patents in fuel cell technology. Mr. Forrester testified that the EC/IR II
    used both electrode chemical, also known as fuel cell, technology and infrared
    technology. The electrode chemical sensor, or fuel cell, is used to “quantitate the
    ethanol” while the infrared technology is used to monitor the breath sample as it is given
    to ensure that the concentration of alcohol increases from the beginning of the sample to
    the end. Use of the infrared technology in this manner eliminates high breath alcohol
    readings due to the presence of alcohol in the mouth. He explained fuel cell technology:
    [A] fuel cell . . . is a porous, plastic disc about the size of a
    quarter, and on either side of that disk we have a pressed
    platinum black, which is a catalyst, and that disk is
    impregnated with an electrode light, it’s an acid . . . and this
    whole thing is encased, and what happens is when you draw a
    sample across the cell, if there’s alcohol in that sample, the
    alcohol becomes an energy source, the alcohol is oxidized on
    the fuel cell surface, hydro genions are created that go
    through [the] porous disc to the other side of the cell, negative
    charge on one side of the cell or a positive charge on the other
    side of the cell, and . . . if you link those two, you’ll have
    electrons flow out of the cell from one side to the other, and
    that is current, which is what we’re measuring. And what’s
    sort of unique and neat about a fuel cell is that the amount of
    alcohol that you get on the cell is proportionate to the current,
    and that’s why a fuel cell is linear, because essentially what
    you know is that if you put X amount of alcohol in the cell
    and you get Y current, if you put two X on, you get two Y
    current. You put [four] X on, you get four Y current. If you
    put zero alcohol on, you get zero current.
    2
    Mr. Forrester emphasized that the machines were physically manufactured by a subcontractor
    called Alca Tech to the specifications set by Intox.
    -4-
    He said that a non-linear reading would indicate the creation of energy, which is
    “physically impossible” using fuel cell technology. Mr. Forrester testified that “a
    tremendous amount of testing” and “a tremendous amount of anecdotal information”
    confirmed that fuel cell-based instruments are linear. He said that Intox had 300,000
    instruments in the field running more than 25 million tests a year and that he was “not
    aware of a single fuel cell that has produced a nonlinear result at any age over” the 20
    years that Intox had been manufacturing fuel cell instruments. Mr. Forrester testified that
    it was the need for linearity that led Intox to switch from infrared technology to fuel cell
    technology. He said that while age might lessen the fuel cell’s ability to quantify the
    amount of alcohol, causing it to provide a lower than actual reading, age would not affect
    the machine’s linearity.
    Doctor Jimmy Valentine testified as an expert in analytical chemistry,
    chemistry, organic chemistry, petrology, spectrometry, thermodynamics, and infrared-
    based and electric-fuel-cell-based breath testing. He said that multi-point calibration was
    the generally accepted scientific method for analytical testing because it “shows that a
    given instrument is responding in a linear fashion over what we call a dynamic range.
    Any machine that has any type of detector that’s used for analytical testing will have a
    point in which it no longer responds linear.” Doctor Valentine opined that the data that
    he had reviewed, which was limited solely to the pleadings filed by the parties, did not
    show that the EC/IR II could establish linear dynamic range, explaining, “What’s being
    done here in the State of Tennessee . . . is that a single-point is used to call that
    calibration on this machine, and that simply goes against all analytical principles that you
    would attempt to calibrate any machine with a single point.” He emphasized that “the
    only scientific practice that’s accepted is multi-point calibration.” He said that it would
    be preferable to calibrate the machine using three points that did not include .08 and then,
    after calibration, test the machine using a known .08 standard purchased from the
    National Bureau of Standards. Doctor Valentine also questioned the use of dry-gas
    standards for testing, noting that it was “a fairly poor simulator of the human lung,” and
    suggested that the machines should be calibrated every day.
    Doctor Valentine testified that in preparation for his testimony, he
    researched a database of peer-reviewed articles and was unable to find a single article on
    linearity in fuel cell technology. Doctor Valentine acknowledged that he had also never
    encountered a fuel cell that became non-linear over time. Additionally, he noted that no
    breath testing machine need be tested beyond the upper limits of human alcohol
    tolerance. He said that quality control testing performed by the manufacturer established
    that the machines were linear when they left the company.
    Michael G. Link, who was declared an expert in “repair, calibration[,] and .
    . . follow-up testing,” testified that single-point calibration was not an accepted practice
    -5-
    in the scientific community or in the field of regulatory testing. He said that if the EC/IR
    II is not capable of multi-point calibration, it should be subjected to multi-point accuracy
    checks. He noted that, if he “were designing a test protocol,” he would test the machine
    for accuracy at the breath alcohol levels that trigger criminal violations. He stated that
    fuel cells contained electronics that were “very, very touchy,” and they should be tested
    regularly to ensure that they are not being affected by being inside the squad cars. He
    said that the machines should be calibrated and subjected to a multi-point accuracy check
    any time they are sent for repairs or subjected to a software change. Mr. Link said that he
    had observed non-linearity in a fuel cell instrument, but he had no familiarity with any
    breath testing machine.
    At the conclusion of the hearing on the defendants’ motion, the trial court
    took the motion under advisement. In its written order denying the defendants’ motion,
    the trial court provided a thorough statement of facts but made no actual findings of facts.
    Citing State v. Cook, 
    9 S.W.3d 98
    , 103 (Tenn. 1999), the court observed generally that
    “pretrial motions to suppress evidence address themselves to the preclusion of evidence
    illegally seized, whereas trial objections are intended to address whether the prosecution
    has laid the proper foundation for the admission of the evidence,” and, citing State v.
    Conway, 
    77 S.W.3d 213
    , 219 (Tenn. Crim. App. 2001), observed specifically that the
    court should examine “only . . . the evidence presented at trial to determine if the
    foundational requirements of State v. Sensing, 
    843 S.W.2d 412
    (Tenn. 1992), had been
    met.” Although the trial court denied the motion, the court did not rule that the State
    would be permitted to use the breath test results at trial. Instead, the court determined
    that “the breath test results in all of these cases is admissible, assuming that the threshold
    requirements of Sensing are met at trial, notwithstanding the possible merits of the
    defendants’ contentions.” (emphasis added). Finally, the court concluded that “the
    evidentiary foundation established in Sensing remains the standard for the admissibility
    of breath test results with the trier of fact assigned the responsibility to determine the
    weight to be given to such evidence.”
    Following the trial court’s denial of the motion to suppress, the defendant
    moved the court to exclude the breath test results as unreliable under the rules of
    evidence. The defendant argued that Sensing did not control admissibility of the breath
    test results in this case because the machine at issue in Sensing was not the EC/IR II and
    because Sensing predated the rulings in Daubert v. Merrell Daw Pharmaceuticals, Inc.,
    
    509 U.S. 579
    (1993), and McDaniel v. CSX Transportation, Inc., 
    955 S.W.2d 257
    (Tenn.
    1997), which dealt with the admissibility of scientific evidence under Evidence Rule 702.
    The trial court denied the motion, noting that in State v. Edison, 
    9 S.W.3d 75
    (Tenn.
    1999), which post-dated both Daubert and McDaniel, our supreme court concluded that if
    the State can establish the Sensing prerequisites by a preponderance of the evidence, a
    -6-
    trial court “should admit” the results of breath alcohol testing without making any
    findings under Evidence Rule 702.
    After the trial court denied that motion, the defendant moved the trial court
    to allow an interlocutory appeal pursuant to Rule 9 of the Tennessee Rules of Appellate
    Procedure. The trial court denied the motion. Thereafter, the defendant’s case, along
    with the other consolidated DUI cases, were transferred from Judge Mark Fishburn to
    Judge Amanda McClendon.
    On August 21, 2014, the defendant pleaded guilty to a charge of DUI per
    se. Pursuant to the plea agreement between the parties, the State dismissed the alternative
    DUI charge and a charge of resisting arrest, and the defendant reserved the following
    certified questions of law that were incorporated by reference into the judgment:
    (1). Whether the trial court erred in refusing to suppress the
    Defendant’s BAC, pursuant to the requirement mandated in
    State v. Sensing, 
    843 S.W.2d 412
    (Tenn. 1992), that “the
    evidentiary breath testing instrument was certified by the
    forensic services division, was tested regularly for accuracy
    and was working properly when the breath test was
    performed.” The State contends that the standards and
    procedures used by the T.B.I. in this matter fully complied
    with the Sensing requirements. The Defense contends that
    the standards and procedures used by the T.B.I failed the
    Sensing requirements because the Defense contends that the
    T.B.I. failed to certify the scientific linearity of the EC-IR II,
    and therefore failed to properly certify the EC-IR II machine
    in question and failed to properly test the machine for
    accuracy at the time in which it was used.
    (2). Whether the trial court erred in refusing to suppress the
    Defendant’s BAC pursuant to its gate-keeping function to
    preclude the admission of unreliable evidence.
    The agreed order specifies that the parties and the court agreed that the questions are
    dispositive of the case.
    As in any other appeal before this court, our first concern is whether this
    court is authorized to hear the case. Jurisdiction to hear a direct appeal following a guilty
    plea generally must be predicated upon the provisions for reserving a certified question of
    -7-
    law. Tennessee Rule of Criminal Procedure 37(b), which, as is applicable in this case,
    provides that a defendant
    may appeal from any judgment of conviction . . . on a plea of
    guilty . . . if . . . the defendant entered into a plea agreement
    under Rule 11(c) but explicitly reserve[s]‒with the consent of
    the state and of the court‒the right to appeal a certified
    question of law that is dispositive of the case.
    Tenn. R. Crim. P. 37(b)(2)(A). To perfect an appeal of a certified question under these
    circumstances, the following requirements must be met:
    (i) the judgment of conviction or order reserving the certified
    question that is filed before the notice of appeal is filed
    contains a statement of the certified question of law that the
    defendant reserved for appellate review;
    (ii) the question of law as stated in the judgment or order
    reserving the certified question identifies clearly the scope
    and limits of the legal issue reserved;
    (iii) the judgment or order reserving the certified question
    reflects that the certified question was expressly reserved with
    the consent of the state and the trial court; and
    (iv) the judgment or order reserving the certified question
    reflects that the defendant, the state, and the trial court are of
    the opinion that the certified question is dispositive of the
    case[.]
    Tenn. R. Crim. P. 37(b)(2)(A)(i)-(iv). In light of the dispensatory nature of a certified
    question appeal, our supreme court firmly rejected a rule of substantial compliance, see
    State v. Armstrong, 
    126 S.W.3d 908
    , 912 (Tenn. 2003), and instead demanded strict
    adherence to Rule 37(b), see State v. Pendergrass, 
    937 S.W.2d 834
    , 836-37 (Tenn. 1996).
    Because a basic requirement for a certified question appeal is that the question actually be
    dispositive of the case, “the reviewing court must make an independent determination
    that the certified question is dispositive.” State v. Dailey, 
    235 S.W.3d 131
    , 135 (Tenn.
    2007).
    In this case, although the trial court denied the defendant’s motions, the
    court did not rule that the breath test results would be admissible at the defendant’s trial.
    -8-
    Instead, the court deferred any ruling on the admissibility of the evidence, concluding
    that, so long as the State could establish the Sensing prerequisites at trial by a
    preponderance of the evidence, the breath test results would be admissible. Indeed, the
    court noted serious misgivings about the State’s ability to satisfy the Sensing
    requirements in light of the evidence offered at the hearing but nevertheless concluded
    that the issue was one to be addressed at trial. Because the court’s ruling neither admitted
    nor excluded evidence, it was not a final determination of the merits of the defendant’s
    claim, and, as such is not dispositive. The court’s conclusion that Sensing governed the
    admissibility of the breath test results was not dispositive of the case based on established
    case law.
    Similarly, the court’s refusal to suppress the breath test results prior to trial
    in its role as evidentiary gatekeeper was not dispositive. The court reiterated that Sensing
    governs the admissibility of the breath test results and that, if the State could satisfy the
    Sensing requirements, the results would be admissible. If the State so chose, it could
    pursue admission of the evidence via Evidence Rules 701 and 702. That method of
    introduction would require the presentation of expert testimony. As this court has
    previously explained,
    Our view is that if the [S]tate complies with the requirements
    of Sensing, it is entitled to the presumption that the test results
    are reliable and the results may be admitted into evidence
    without the benefit of an expert. If not, the [S]tate may still
    use traditional rules of evidence to lay the foundation for
    admitting the evidence but there is no presumption of
    reliability.
    State v. Deloit, 
    964 S.W.2d 909
    , 913 (Tenn. Crim. App. 1997).
    Again, the orders denying the two motions in this case do not contain any
    factual findings and do not constitute a ruling on the merits of the defendant’s challenge
    to the admissibility of the breath test results. In consequence, neither of the certified
    questions is dispositive of the case, and the appeal must be dismissed.
    _________________________________
    JAMES CURWOOD WITT, JR., JUDGE
    -9-