NUMBER 13-07-00090-CV
DORIS THOMPSON, INDIVIDUALLY,
AND TOWNSI FOULKROD
AS LEGAL REPRESENTATIVE OF
KENZEY THOMPSON, MINOR,
INDIVIDUALLY AND ON BEHALF
OF THE ESTATE OF MICHAELYNN THOMPSON, Appellees.
MEMORANDUM OPINION
Before Justices Yañez, Benavides, and Vela
Memorandum Opinion by Justice Yañez
Alza Corp. ("Alza") appeals a judgment on a jury verdict rendered against it in a wrongful death case involving allegations of negligence, strict products liability manufacturing defects, (2) and gross negligence regarding its manufacture of the Duragesic patch. The manufacturing defect asserted is a flaw in the seal of the Duragesic patch which allowed decedent, Michaelynn Thompson, to receive a fatal overdose of fentanyl, a potent narcotic. (3) The negligence and gross negligence claims include complaints that Alza continued to employ a visual inspection process, intended to spot manufacturing defects in the Duragesic patches as they passed on a conveyor belt, even after Alza knew that such a process was inadequate to detect the flawed patches in order to remove them from distribution to patients. (4)
The plaintiffs-appellees ("the Thompson family") (5) tried their claims arising from the death of Michaelynn Thompson against Alza before a jury in the 113th Judicial District Court in Harris County, Texas, with the Honorable Patricia Hancock presiding. The jury found that there was a manufacturing defect in the Duragesic patch worn by Thompson which was a cause of her death, and further found that Alza was negligent, but declined to find Alza grossly negligent. The trial court rendered judgment on the jury's verdict. Alza appeals by five issues and multiple sub-issues, generally challenging the sufficiency of the evidence and two evidentiary rulings. We affirm.
The Duragesic patch is a prescription pain patch utilized to treat moderate to severe chronic pain. It is a transdermal system that adheres to a patient's skin and is designed to provide the patient with a continuous, systematic delivery of fentanyl, a potent opioid analgesic, for a period of seventy-two hours. This patch is comprised of four layers: an external backing of polyester film, a drug reservoir containing fentanyl and alcohol in a gel solution, a membrane that controls the rate of fentanyl delivery to the skin surface, and an adhesive worn next to the skin. The Duragesic patches come in several strengths.
The Duragesic 75 mcg/hr patch worn by Thompson at the time of her death (6) was manufactured on Alza's Bodolay machine. The patches at issue were made at a rate of about four per second, and the quality control process for the patches entailed a visual assessment by two inspectors as the patches went by on a high-speed conveyor belt. From 2001 through 2004, Alza documented many instances where patients had received Duragesic patches manufactured on its Bodolay machine with flawed seals (7) that were not detected by Alza's visual inspection process. The number of patient complaints regarding defective 75 mcg/hr patches escalated sharply in 2003 and early 2004, ultimately involving an inspection by the United States Food and Drug Administration ("FDA") and the recalls of four lots of 75 mcg/hr Duragesic patches.
Rudolph Holland, a senior technician on the Bodolay running the Duragesic line for Alza, testified about the specific manufacturing process and product components that constitute the Duragesic patch. At trial, Holland testified that he was not mechanically inclined, and that no one ever asked him what could have contributed to the number of leaking patches.
Holland testified that he was aware that Alza had been notified of a fold-over defect in a Duragesic patch in 2001. Holland's sworn testimony established that in 2001, after learning that a fold-over occurred, and after realizing how the defect occurred during its manufacturing process, the company made no changes to its quality control procedures; however, Alza did instruct its employees to look for fold-over defects during the visual inspection process. Holland testified that Alza's visual inspection process entailed watching 120 lines of product go by per minute. Holland testified that he has never seen a fold-over defect. Alza's visual inspections detected no problems with the two-and-a-half million Duragesic patches that were ultimately recalled by Alza in 2004 because of the prevalence of fold-over defects.
In 2003, Holland "thought everything was fine" regarding the company's manufacturing process, and thought that the visual inspection process was sufficient, but conceded in retrospect that "we could have made some changes," and concluded in his testimony before the jury that asking the employees to look harder for defects was not sufficient to prevent product defects. Holland testified that Alza improved and modified some of its inspection procedures after the Duragesic recall in 2004, so that the Duragesic patches now go through a weight check, a pressure check, a visual inspection, and quali-pack testing with a vacuum. He further testified that quality assurance employees also hold "line audits" every eight hours.
Holland did not know how many leaking patches Alza produced; however, based on patient complaints to the company, 35 out of 40 lots of 75 mcg patches manufactured between January 1, 2003 and June 8, 2004 contained leaking patches. According to documents he reviewed at trial, Holland testified before the jury that reports of leaking patches were lower in two of the five recalled batches than in other batches that Alza chose not to recall. Holland expressly admitted in testimony to the jury that he would have notified the FDA of the escalating number of fold-over defects earlier than Alza did.
As mentioned above, when complaints from the public about the Duragesic patch escalated sharply in late 2003 and 2004, the FDA investigated this problem. According to the FDA's inspection report, (8) Alza characterized the initial problem as an isolated incident, and Alza's approach to the problem was to re-train the operators:
The firm first came across a fold over defect from a field sample return back in 10/01 . . . and assigned a probable cause as the polyester/EVA film being folded over during a roll change or start up. The firm's corrective action was to train operators to "heightened awareness" and that the fold over was an "isolated incident." This corrective action did not prevent the fold over defect from reoccurring . . . .
The FDA's inspection report found Alza's remedial actions inadequate:
The FDA's inspection report also specifically addressed Alza's training of the operators:
[T]his approach of reactionary training of operators from field complaints did not appear to alleviate these types of product defects due to this corrective action being seen in one Exception Report after another. I expressed the possibility of the firm to institute a preventative training program for the operators in addition to their routine GMP training on a regular basis through the year. All parties agreed that this was a positive approach and provided a memo reflecting this preventative training program . . . .
The FDA inspection report also noted that the "defect would be incorporated into the patch and could go undetected with current QA checks in place." Similarly, the FDA proposed ways "to prevent this defect from reoccurring," and Alza agreed that "equipment changes and controls are the best remedies for preventing this defect in the future." In this context, the FDA report documented numerous changes in Alza's Bodolay machine, which was at the end of its usefulness and scheduled for replacement, in order to reduce the incidence of manufacturing defects. The report includes an evaluation of this process with Alza's management:
On 6/08/04, [the FDA representative] held a discussion with [Alza representatives] regarding the Bodolay I FFS and [its] ability to produce quality product or should the machine be retired. All three expressed their desire to improve operations and that the implementation of the HH FFS is on top priority to be validated and put into production to be the main manufacturing line for Duragesic and therefore reserve the Bodolay I FFS as a backup machine. They continued to express that the changes made to the Bodolay I FFS and the implementation of the vision system on the Bodolay I should and will improve the quality of product produced.
As a result of this defect, Alza recalled about 2,500,000 of its Duragesic patches in 2004. (9) Also during 2004, Dr. Elias Benhamou, a pain management specialist who is board-certified in anesthesiology, prescribed Duragesic patches for Michaelynn Thompson, who was suffering chronic back pain after a car crash. (10) Dr. Benhamou prescribed these Duragesic patches during the time period after Alza began its defect investigation but before Alza initiated the recall. (11) At trial, Dr. Benhamou testified that an overdose of fentanyl results in symptoms including discomfort, difficulty breathing, sweating, dizziness, and nausea, and that a Duragesic patch leaking an uncontrolled amount of fentanyl could result in the death of the patient.
During the early morning hours of February 13, 2004, Thompson began vomiting
and complaining of dizziness and back pain. The vomiting continued on an hourly basis
until morning. Thompson's thirteen-year old daughter, Kenzey Marie Thompson, ultimately
called an ambulance for her at approximately 9:00 a.m. While transporting Thompson to
the emergency room, the emergency technicians performed an electrocardiogram
("EKG") (12)
on Thompson showing normal results. Thompson's medical records from Fort Bend Methodist Hospital state that
Thompson was wearing a "Duragesic patch" upon admission to the emergency room.
These records indicate that Thompson's complaints included dizziness, nausea, vomiting,
abdominal pain, and left flank pain. Nurse Camille Green testified that Thompson
complained of abdominal pain and dizziness. Kenzey testified that her mother was acting
"weird" at the hospital. Dr. Donnie Evans, the emergency room physician, treated Thompson. He ordered
numerous tests, including a chest pain evaluation, an electrocardiogram, and cardiac
enzyme tests. The tests showed no arrhythmia or heart damage. After Thompson's
complaints of pain subsided, Fort Bend began preparations to release her. Kenzey
telephoned members of their church to come pick them up from the hospital, and
Thompson and Kenzey waited to be released. Kenzey testified that her mother told her
that she was really tired and that she was going to take a nap. Kenzey slept also. Upon
waking, Kenzey attempted to wake her mother but was unable to do so. She informed a
nurse, who told her that her mother would have to be awake to be discharged. Kenzey
saw her mother, apparently asleep, lying on her stomach. Kenzey noted that her mother's
legs were bluish-purple with goose bumps, and she was lying face-down in her pillow with
her hands up by her face, and her fingernails were purple. Kenzey attempted to rouse her
mother because the nurse had said they could leave if she could wake her, but Kenzey
was unable to do so. Kenzey testified that she had seen a patch on her mother's abdomen the night
before, and she said her mother was really good about going to her doctor and changing
out her patches. During this period, Nurse Karen Marie Cothran testified that Kenzey approached her
sometime before 3:50 p.m., and told her that her mother was sleeping a lot. Cothran told
Kenzey that her mother would have to wake up before she could be discharged. Cothran
entered the room at 3:50 p.m. and found Thompson, who was not breathing, laying on her
stomach with her face in the pillow. She moved Thompson and sounded an emergency
code. Dr. Evans testified that when he entered the room, Thompson's back was noticeably
blue. He and a team began resuscitation efforts to no avail. He saw the patch on
Thompson's back and ripped it off because it was "an unknown," and threw it away. On
the same day that Thompson died, Alza initiated the initial phase of the recall, and several
days later, Dr. Evans received notice of the recall. He promptly contacted Thompson's
family so that they could look into whether the patch had any causal relationship to her
death. Dr. Evans testified that he had no concerns about drug toxicity, nor did he see signs
thereof when he was treating her, but at the same time, he did not realize she had the
patch on. In this regard, we note that testimony at trial established that symptoms of an
overdose of fentanyl include, inter alia, nausea, vomiting, sleepiness, dizziness, and
respiratory depression. As a result of the contact between Dr. Evans and the Thompson family, the family
repeatedly requested the medical examiner to test the level of fentanyl in Michaelynn's
blood. Despite these requests, the medical examiner performed a routine medical drug
screen, with negative results, but did not test Thompson's blood for fentanyl. Evidence at
trial confirmed that it is not possible to determine whether someone has a toxic level of
fentanyl in their blood without testing for it. The medical examiner did save a femoral
blood sample and sent it to an outside laboratory, Accu-Chem, for testing at the Thompson
family's request. This test showed that Thompson had a blood fentanyl quantitation of 11.4
ng/ml. A therapeutic range for fentanyl is 1.0 to 2.0 ng/ml, and a toxic range is 3.0 to 5.0
ng/ml. Stated otherwise, the test confirmed that Thompson's blood showed a toxic level
of fentanyl that was over ten times the low end of the expected range based on her
fentanyl prescription, and over five times the uppermost end of that range. (13) Having not performed a test for fentanyl, the Harris County Medical Examiner's
report showed Thompson's cause of death as "cardiomegaly with biventricular hypertrophy
associated with obesity," which, according to the testimony at trial, simply means that
Thompson had an enlarged heart. According to the autopsy report, Thompson was forty-two years old, six feet tall, and weighed 367 pounds. At the trial arising from Thompson's death, Mika Greenfield, Alza's corporate
representative and quality assurance manager, testified regarding Alza's knowledge of
product defects and its manufacturing process. She conceded that fold-over defects exist
where an entire side of the patch is leaking, and she agreed that fold-over defects and
stringer leakers tend to release fentanyl in an uncontrolled way. Despite the fact that she
was Alza's corporate representative, her testimony was conflicting regarding her knowledge
of whether fentanyl was a dangerous drug, and she was unaware whether morphine was
more potent than fentanyl. Greenfield testified Alza began receiving increasing numbers of public reports of
leaking patches as early as January 7, 2004. She testified that, for instance, Alza
confirmed forty-two fold-over defects in one of its lots of Duragesic patches. She admitted
that these defects were originally found by the public, and not by Alza or its employees.
She was unable to describe the safety process utilized by Alza to find product defects in
2003, and conceded that Alza's technicians found no fold-over defects in any of the
Duragesic patches produced during the last six months of 2003. Greenfield confirmed that Alza was notified of more than one fold-over defect in
Duragesic patches in 2001. At that time, the company's response was to heighten visual
inspection of the patches in a two-foot space on a conveyer belt as the product was
traveling through the Bodolay system. Greenfield testified that she did not know what changes were made on the Bodolay
after the product recall to ensure that it was not producing defective products even though,
as part of the company's quality assurance team, her responsibility was to ensure that the
manufacturing process is completed without defects. She did offer sworn testimony,
however, admitting that all changes that were put in place after the product recall, including
a mechanical visual system, different evaluating methods, equipment modifications, and
equipment cleaning, would have been available for use in 2001. Greenfield testified that Alza did not examine the recalled patches to determine how
many of the recalled patches actually had defective seals or fold-over defects. She
conceded that after determining that there was a fold-over defect problem, Alza released
additional patches for public distribution without any additional inspection of the patches.
Greenfield further testified that internal correspondence at Alza acknowledged the fact that
it was keeping the Duragesic patches on the market, despite the defects, because of
product shortages. Finally, she admitted that the FDA's inspection processes resulted in
Alza's making changes to improve its processes and equipment. Juanita Hawkins, the worldwide vice president of Quality Assurance and Technical
Operations for Cordis Corporation, which is a subsidiary of Johnson & Johnson, (14) also
provided testimony on behalf of Alza. In 1995, Hawkins worked for Janssen as the vice
president of quality assurance. Hawkins testified that Alza would not and should not
knowingly release a lot that contained defective products; nevertheless, she admitted that
the FDA concluded that, and internal documents confirmed that, the company continued
to release products for commercial distribution despite knowing of product defects therein. Dr. David Upmalis also testified on behalf of Alza. He is a general practice
physician who joined the pharmaceutical industry, and now works with Pharmaceutical
Research and Development ("PRD"), a research group that "supports Alza, among other
companies." He does clinical research in painkillers, "providing medical support for
Duragesic and a couple of other pain killers that are in the marketplace." He handles
correspondence with regulatory agencies, including the FDA, regarding product issues and
works to obtain approval for drugs. He worked with pediatric clinical trials for the Duragesic
patch and helped derive labeling for the patch. Dr. Upmalis testified about two general types of leaks that a Duragesic patch could
exhibit. A stringer leaker is a patch that has a piece of gel stuck in the seal. According to
Dr. Upmalis, Alza's assessments concluded that there was no significant health risk posed
by stringer leakers. Dr. Upmalis also testified about fold-over defects, where the seal
breach covers an entire side of the patch. Dr. Upmalis testified that, in response to the escalating number of public complaints
about product defects, Alza formed a health hazard assessment committee which was
asked to look at fold-over defects on President's Day weekend, February 14-15, 2004.
According to Dr. Upmalis, the health risk involved in fold-over defects was higher than for
stringer leakers because more fentanyl gel could leak from a fold-over defect. Dr. Upmalis
testified that each lot of Duragesic patches contained around 400,000 units, and Alza
ultimately estimated that approximately 3,600 patches would be affected by fold-over
defects in each lot. Dr. Upmalis testified that Alza utilized Medline searches and three
studies, which were not peer reviewed, in its determination regarding the health risks
incident to leaking Duragesic patches. In its research, Alza found many incidents of
adverse events related to the rapid release of fentanyl. Dr. Upmalis was involved in writing the Physician's Desk Reference ("PDR")
materials on the Duragesic patch. The warning in the PDR states: "Duragesic patches are
intended for transdermal use on intact skin only. Using damaged or cut Duragesic patches
can lead to the rapid release of the contents of the Duragesic patch and absorption of a
potentially fatal dose of Fentanyl." Dr. Upmalis testified that this warning resulted from
people's "abuse" of, or damage to, the patches. Despite the explicit text of the PDR
warning, Dr. Upmalis testified that the warning of adverse health consequences therein
concerns the absorption of fentanyl through mucous membranes rather than skin. Dr. Upmalis explained that the adverse drug reporting system requires doctors to
report to the company and to the FDA if there is an adverse drug event. He admitted that
under-reporting is a known problem, and he has heard that only one to ten percent of
adverse reactions are reported. He agreed that the FDA thought that under-reporting is
particularly high with the Duragesic patch because of the nature of the patient population
utilizing the patch. Dr. Upmalis examined Duragesic adverse event reports, which
included, in 2002, for instance, 53 cases of hypoventilation, 270 cases of dyspnea
(shortness of breath), 15 cases of respiratory arrest, 33 cases of respiratory depression,
8 cases of respiratory disorder not otherwise specified, 8 case of respiratory distress, and
14 cases of respiratory failure. The 2003 adverse event report numbers were similar.
According to Dr. Upmalis, given the incidence of under-reporting, these adverse reactions
were just "the tip of the iceberg." (15) Dr. Leslie Benet, a pharmacologist who is a professor at the University of California,
provided further testimony on Alza's behalf. Dr. Benet testified that he works for almost
all drug companies. While Alza does not pay him directly for his work, Alza pays the
university for which he works, which in turn funds his laboratory. Dr. Benet conceded that
the FDA did not concur with the pharmaceutical-industry funded tests which downplayed
the health risks from Duragesic patch defects, and that the FDA had determined that the
justification presented to support the safety of leaking patches was inadequate. (16) Dr. Benet admitted that Alza's Duragesic labels do not corroborate his trial testimony
regarding adverse health risks associated with damaged patches. He attempted to explain
this discrepancy as a result of the fact that all labeling is supposed to include all potential
toxic eventualities whether they occur or not: "And today, because of litigation such as this,
every company lists everything that could potentially lead to a litigious situation." Dr. Benet
conceded that Alza knew about a number of leaking patches before 2005, and at that time,
Alza did not place a warning on the Duragesic packaging. According to Dr. Benet, the
warning that Alza ultimately included on the Duragesic labels is directed to patient
caregivers, and not to the patch-users themselves, because the caregivers are more likely
to have a toxic reaction to exposure. Upon cross-examination, Dr. Benet admitted that the
Duragesic label states that it is for use only in opioid tolerant patients, and that because
serious or life-threatening hypoventilation could occur, the Duragesic transdermal system
is contraindicated in patients who are not opioid-tolerant. The label further states that
damaged or cut patches can lead to rapid release of the contents and absorption of a fatal
dose. Dr. Benet ultimately concluded that Alza and Janssen both knew that fentanyl was
a toxic substance and the exposure of fentanyl at an uncontrolled rate might be dangerous;
however, Dr. Benet testified that in this case, given that Thompson had been on pain
medication for a number of years and had been wearing a patch for more than a month,
he did not believe that fentanyl gel leaking out on her skin would pose a health risk to a
reasonable degree of scientific certainty. As mentioned above, the Thompson family's case was tried before Judge Hancock
in Harris County, Texas, and the jury returned affirmative findings concerning the
manufacturing defect and Alza's negligence, but did not find Alza grossly negligent. Alza
challenges Judge Hancock's judgment on the verdict by five issues. Alza's first three
issues concern the sufficiency of the evidence supporting the verdict. In reviewing the legal sufficiency of the evidence, we view the evidence in the light
most favorable to the jury's findings, giving full credit to all favorable evidence if any
reasonable person could, and disregarding contrary evidence unless reasonable persons
could not. City of Keller v. Wilson, 168 S.W.3d 802, 807 (Tex. 2005). In Texas, the jury
is the sole judge of the witnesses' credibility, and that time-honored principle is a part of
our standard of review: Jurors are the sole judges of the credibility of the witnesses and the weight
to give their testimony. They may choose to believe one witness and
disbelieve another. Reviewing courts cannot impose their own opinions to
the contrary. Most credibility questions are implicit rather than explicit in a
jury's verdict. Thus, reviewing courts must assume jurors decided all of them
in favor of the verdict if reasonable human beings could do so. Courts
reviewing all the evidence in a light favorable to the verdict thus assume that
jurors credited testimony favorable to the verdict and disbelieved testimony
contrary to it. Id. at 819 (footnotes and paragraph break omitted). Accordingly, we must deny a legal
sufficiency challenge unless: (1) there is a complete absence of evidence of a vital fact;
(2) all the evidence offered to prove a vital fact amounts to only a mere scintilla and no
more; (3) the rules of law or evidence completely forbid any consideration of the only
evidence offered to prove a vital fact; or (4) the evidence which the jury was compelled to
consider and believe conclusively establishes the opposite of the vital fact. See id. at 810
& n.15-16. These vital facts may be proved by direct or circumstantial evidence. See Ford
Motor Co. v. Ridgway, 135 S.W.3d 598, 601 (Tex. 2004); Russell v. Russell, 865 S.W.2d
929, 933 (Tex. 1993). In reviewing the factual sufficiency of the evidence, we must consider and weigh all
the evidence, and we should set aside the judgment only if it is so contrary to the
overwhelming weight of the evidence as to be clearly wrong and manifestly unjust. Pool
v. Ford Motor Co., 715 S.W.2d 629, 635 (Tex. 1986); Cain v. Bain, 709 S.W.2d 175, 176
(Tex. 1986). By its first and third issues, Alza challenges the legal and factual sufficiency of the
evidence supporting the jury's finding of a manufacturing defect which was a producing
cause of Thompson's death. In the instant case, the jury found that "there was a
manufacturing defect in a Duragesic patch worn by Michaelynn Thompson at the time the
patch left the possession of Alza that was a producing cause of Michaelynn Thompson's
death." In its first issue, Alza contends that there is legally and factually insufficient evidence
to support the jury finding that a manufacturing defect, existing at the time of manufacture,
caused Thompson's death. In connection with this issue, Alza argues: (1) there was no
proof of proper use of any Duragesic patch in question, and specifically, no proof regarding
a 75 mcg/h patch; (2) there was no evidence to support a finding of malfunction; and (3)
the evidence conclusively established that Thompson could not have been wearing a 75
mcg/h patch from a recalled lot when she died. "A manufacturing defect exists when a product deviates, in its construction or
quality, from the specifications or planned output in a manner that renders it unreasonably
dangerous." Cooper Tire & Rubber Co. v. Mendez, 204 S.W.3d 797, 800 (Tex. 2006); see Bic Pen Corp. v. Carter, 251 S.W.3d 500, 509 (Tex. 2008). A plaintiff must prove that the
product was defective when it left the hands of the manufacturer and that the defect was
a producing cause of the plaintiff's injuries. Ridgway, 135 S.W.3d at 600; Torrington Co.
v. Stutzman, 46 S.W.3d 829, 844 (Tex. 2000). A specific defect "must be identified by
competent evidence and other possible causes must be ruled out." Ford Motor Co. v.
Ledesma, 242 S.W.3d 32, 42 (Tex. 2007) (quoting Nissan Motor Co. v. Armstrong, 145
S.W.3d 131, 137 (Tex. 2004)). Both direct and circumstantial evidence may be used to establish any material fact.
Lozano v. Lozano, 52 S.W.3d 141, 149 (Tex. 2001); Browning-Ferris, Inc. v. Reyna, 865
S.W.2d 925, 928 (Tex. 1993). To raise a genuine issue of material fact, however, the
evidence must transcend mere suspicion. Ridgway, 135 S.W.3d at 601. Evidence that
is so slight as to make any inference a guess is in legal effect no evidence. Id.; Lozano,
52 S.W.3d at 148. Alza contends that appellees' claims are dependent on "proof of proper use of the
patch," citing Parsons v. Ford Motor Co., 85 S.W.3d 323, 329-30 (Tex. App.-Austin 2003,
pet. denied) (citing Plas-Tex, Inc. v. U.S. Steel Corp., 772 S.W.2d 442, 444-45 (Tex.
1989)). Alza contends that appellees presented no evidence of how any Duragesic patch
was handled by Thompson on the occasion in question. According to Alza, "[t]he patch
that was removed by Dr. Evans from [Thompson's] back was reportedly unremarkable.
The strength of the patch found on her body at the time of her death was never
established. The decedent's daughter claimed that Thompson had a patch on her
stomach on the day she died." Manufacturers face liability only for products which are defective when used in the
intended manner or when used in a reasonably foreseeable way. See Houston Lighting
& Power v. Reynolds, 765 S.W.2d 784, 786-87 (Tex. 1998) (liability is "limited to uses that
are objectively reasonable to expect . . . it does not encompass uses . . . which represent
wholly unexpected product misuse.") (citations omitted). (17) In this regard, a manufacturing
defect exists if a product does not conform to the design standards and blueprints of the
manufacturer, and the flaw makes the product more dangerous and therefore unfit for its
intended or reasonably foreseeable uses. See Benavides v. Cushman, Inc., 189 S.W.3d
875, 881 (Tex. App.-Houston [1st Dist.] 2006, no pet.). Dr. Benhamou, the physician who treated Thompson for pain management and who
prescribed Duragesic for her, reviewed Thompson's medical records and testified
regarding her usage of the Duragesic patches. Thompson had worn Duragesic patches
ten separate times, from January 13 until February 11, without incident or adverse effect. (18)
The timing of her prescription refills for Duragesic patches fit with the prescribed usage of
the product, and the timing would also illustrate that Thompson was wearing one of the
newly prescribed 75 mcg patches at the time of her death. Thompson's medical records from the emergency room state that she was admitted
with a "Duragesic Patch." Dr. Evans, the emergency room physician, testified that
Thompson told him she sometimes used the Duragesic patch. He further testified that he
saw the patch on Thompson's back and removed it when attempting to resuscitate her
when she coded because he "thought it may have inhibited our resuscitation [efforts] only
because it was an unknown factor." In short, the evidence indicates that Thompson was
wearing a 75 mcg Duragesic patch at the time of her death. (19) Dr. Benhamou testified that he never had any indication that Thompson was
abusing Duragesic during his course of treating her. He further testified that he never had
any indication that she was not following his instructions with regard to the medication.
Further, Dr. Benhamou testified that he had required Thompson to enter a "medication
contract" with him, entered into evidence at trial, which spelled out Thompson's obligations
with regard to the proper usage of Duragesic. Dr. Robert C. Bux, a physician and board-certified forensic pathologist and medical examiner, who was testifying on behalf of the
Thompson family, concurred that there was "no evidence" that Thompson had misused her
medication. Dr. Vincent DiMaio, a medical examiner testifying on behalf of Alza, conceded
that there was "no physical evidence" that Thompson abused the patches. Based upon the foregoing, we conclude that the jury had before it legally and
factually sufficient evidence from which it may have concluded that Thompson was wearing
and properly using a Duragesic 75 mcg at the time of her death. Alza next contends that there is no evidence to support the jury's finding that the
Duragesic patch malfunctioned. In connection with this issue, Alza argues that the Accu-Chem blood test is unreliable because it was not performed properly, and the 11.4 blood
level reading does not establish a malfunction because of postmortem redistribution. (20)
Alza asserts that Accu-Chem deviated from its protocol in performing its testing and further
failed to perform the testing correctly. Alza also attacks the blood test in its third issue,
contending that: (1) the Accu-Chem test was not performed properly; (2) Accu-Chem did
not validate its methodology; (3) Accu-Chem's methodology of relying on "only one"
calibrator was not shown to be reliable; (4) Accu-Chem's test on Thompson's blood had
technical break-downs; (5) Accu-Chem had insufficient experience testing for fentanyl in
human blood for the results to be considered accurate; and (6) the Accu-Chem test is
based on an inference not supported by the evidence. As stated previously, following Thompson's death, the coroner took blood samples
from Thompson's femoral artery and sent the samples to Accu-Chem for testing.
According to expert testimony at trial, blood from the femoral artery area is not subject to
contamination from gastric contents or trauma and will represent the levels and
concentration of substances in the blood at the time of death. Accu-Chem, an independent
testing agency, analyzed the blood and determined that Thompson had 11.4 ng/ml of
fentanyl in her body. This amount is toxic and is well over three times the therapeutic level
of fentanyl. Accu-Chem utilized gas chromatography/mass spectrometry (GC/MS) technology
to test Thompson's blood. We may consider prior judicial opinions allowing the admission
of evidence in determining whether the underlying theory of scientific evidence is valid.
See Emerson v. State, 880 S.W.2d 759, 767-68 (Tex. Crim. App. 1994); Combs v. State,
6 S.W.3d 319, 322 (Tex. App.-Houston [14th Dist.] 1999, no pet.); Jones v. State, 716
S.W.2d 142, 147 (Tex. App.-Austin 1986, pet. ref'd). Such evidence may be considered
for the first time on appeal. Emerson, 880 S.W.2d at 765. Texas and federal courts have
found the gas chromatography test to be a reliable method for identifying compounds, and
it has been generally accepted in the scientific community. See Combs, 6 S.W.3d at 322;
Wright v. State, 853 S.W.2d 154, 155 (Tex. App.-Corpus Christi 1993, pet. ref'd); Jones,
716 S.W.2d at 145-52; see also United States v. Bynum, 3 F.3d 769, 772 (4th Cir. 1993);
Bolieu v. State, 779 S.W.2d 489, 490 (Tex. App.-Austin 1989, no pet.) (discussing the
GC/MS testing as "the Golden Rule" in the field of toxicology). The scientific theory and
principles underlying GC/MS testing have been proven to be reliable. The experts in the
instant case appeared to agree that the GC/MS procedure has been properly validated and
is considered the most accurate method of identifying and quantifying drug levels. We next turn our attention to whether the methodology of GC/MS was reliably
applied in this specific case. Guadalupe-Blanco River Auth. v. Kraft, 77 S.W.3d 805, 808
(Tex. 2002); Merrell Dow Pharms. v. Havner, 953 S.W.2d 706, 712 (Tex. 1997); In re
S.E.W., 168 S.W.3d 875, 884 (Tex. App.-Dallas 2005, no pet.). In this regard, we note
that the law does not require a "court to admit opinion evidence which is connected to
existing data only by the ipse dixit of the expert." Kerr-McGee Corp. v. Helton, 133 S.W.3d
245, 258 (Tex. 2004) (quoting Gen. Elec. Co. v. Joiner, 522 U.S. 136, 146 (1997)). Dr. Gary Harold Wimbish, a board-certified forensic toxicologist, testified on behalf
of the Thompson family. He has performed more than a thousand GC/MS tests. He
testified that the GC/MS test can be used to determine the level of fentanyl in someone's
body, and it is an "excellent instrument" that is "commonly used" to determine the level of
fentanyl in a sample of human blood. He has used GC/MS on several occasions for that
purpose. He testified that whether a drug level in a person's body is at a fatal level is part
of his field of study. Dr. Wimbish testified that he is familiar with Accu-Chem and its operations. He
testified that Accu-Chem is certified by the American Society of Laboratory Chiefs of
Police, the College of American Pathology, and the Clinical Society for Drug Testing.
Analytical laboratories such as Accu-Chem are accredited by various organizations and,
as part of their accreditation, are regularly inspected and participate in "proficiency studies,"
in which identical portions of the same sample are analyzed by many labs at the same time
in order to check the accuracy of each procedure as it is performed in the individual
participating labs. Accu-Chem participates in proficiency testing by the College of
American Pathology and it is certified by organizations for performing drug testing on
parolees and probationers. Dr. Wimbish testified that he is personally experienced with
Accu-Chem and that it is his "laboratory of choice" for a forensic laboratory. Accu-Chem had specifically passed the College of American Pathology's proficiency
test for fentanyl, although Dr. John Arthur Tarver, the director of Accu-Chem, testified that
Accu-Chem does not frequently perform tests for fentanyl. Dr. Tarver further testified that
an outside organization, the American Society of Crime Laboratory Directors, reviewed and
approved Accu-Chem's testing procedures. Further, the particular procedure used to test
fentanyl was validated externally with a proficiency sample from the College of American
Pathologists. Accu-Chem sent the College a proficiency sample, and with those standards
and control standards, Accu-Chem was within five-tenths standard deviations of the mean
of all the laboratories that performed the test. Dr. Graham Richard Jones, a forensic toxicologist, testifying on behalf of Alza,
acknowledged that Accu-Chem had passed the proficiency test administered by the
College of American Pathologists in March 2004, before Accu-Chem did the test at issue
in this case, and had done "well" on the proficiency test. Accu-Chem's report of October 18, 2004 shows a concentration of fentanyl in
Thompson's blood of 11.4 nanograms per milliliter. (21)
The therapeutic range for fentanyl,
that is, the amount found in the blood related to the relief of pain, is from one to two
nanograms per milliliter. The toxic range for fentanyl, that is, the range that would have
toxicity or adverse and deleterious effects, is three to five nanograms per milliliter. Dr.
Wimbish testified that 11.4 nanograms was a little more than two times the upper limit of
the toxic range for fentanyl, and was toxic enough to be lethal. Specifically, Dr. Wimbish
testified that while Thompson may have developed some tolerance to fentanyl, the tested
level of fentanyl found in her blood was nevertheless three to four times larger than an
amount which she could have tolerated. Based on testimony at trial, the GC/MS test is performed by analyzing and
comparing the unknown blood sample, in this case, Thompson's femoral blood specimen
drawn by the Harris County coroner, with other blood samples where the amount of the
target substance is known. The results for the unknown sample are then compared to the
results for the known samples in order to calculate the amount of the target substance
present in the blood sample. The testimony in this case is complicated by the various
names and descriptions the experts used with regard to these "known samples," of which
we have been able to delineate three basic categories. The first category of known samples discussed by the experts is a set of
"calibrators," which are samples prepared in a matrix of blood, ethanol, or some other
solvent, each of which contains a known amount of the target compound, in this case,
fentanyl. A set of calibration standards includes several samples that contain various
known amounts of the target compound within a specific and broad range. The second set
of known samples involves "ion ratios." The ion ratio is the ratio between the abundance
of two different ions within the same chemical, or two ions representing different chemicals,
such as fentanyl, and the internal standard form of fentanyl. Ion ratios can be used either
qualitatively or quantitatively; in other words, they are employed in either identifying the
substance or in measuring the amount of the substance. The third category of known
samples are quality control samples, which are used to check the reproducibility of results.
A quality control sample is a sample which is prepared in the same type of matrix as the
unknown sample, but which has a known concentration of the target compound that is near
the expected concentration of the unknown sample. Samples from proficiency studies, if
they are sufficiently stable, may be used as external quality controls. In the instant case, Accu-Chem utilized four calibrators and ion ratio tests in its
analysis of the fentanyl quantitation in Thompson's blood. The .5 calibrator failed both the
quantitative error test and the ion ratio test. The 2.0 calibrator passed the quantitative error
test and failed the ion ratio test. The 5.0 calibrator failed the quantitative error test and
passed the ion ratio test. The 20.0 calibrator passed both tests. Utilizing this data, Accu-Chem calculated that Thompson's blood contained 11.4 ng/ml of fentanyl. The result of
these calculations was validated by comparison to the quality control samples, which had
been validated by separate protocols before Thompson's test. Using the quality control
samples, Accu-Chem's analysis of Thompson's blood showed a quantitation of 10.48
ng/ml. Dr. Tarver, the director of Accu-Chem, testified by deposition regarding the test
protocols and results. Dr. Tarver has degrees in biology and chemistry and a master's
degree in biomedical science. He is a diplomat of the American Board of Forensic
Toxicology and a toxicological chemist for the National Registry of Certified Chemists. Dr.
Tarver testified that there are no methodology standards that require a certain number of
calibrators to work to produce a valid test result. While Accu-Chem's standard procedure
required the use of four calibrators, Dr. Tarver rejected two of those calibrators. Dr. Tarver
testified that the rejected controls had no effect on his ultimate conclusions because Accu-Chem validated, through external sources, the actual lot number of the standard calibration
material and the quality control material. Dr. Wimbish reviewed and explained Accu-Chem's written protocols for postmortem
testing for controlled substances. Accu-Chem's standard operating procedures allow Dr.
Tarver, the lab director, to review the analytical protocol and provide that Dr. Tarver has
the final authority on any analytical protocol in the laboratory, particularly with regard to
forensic specimens because of their limited and unique quality. With limited samples, the
laboratory director has to make a decision about reporting results, and often, as per his
instructions in the standard operating procedures, he can modify those within scientific
limits as to the acceptability or reportability of a result under unique circumstances. In
other words, Accu-Chem's standard operating procedures allow Dr. Tarver to deviate from
those procedures. Dr. Wimbish testified that it is not unusual for a laboratory director to
have that kind of discretion with regard to forensic samples. Dr. Wimbish testified that Accu-Chem's test rendered a reliable quantitation within
reasonable scientific certainty. He explained that Accu-Chem derived its results by utilizing
and comparing two separate sample sets, then explained his own calculations verifying
Accu-Chem's results. Accu-Chem used two of the four calibrators to derive its results. Dr.
Wimbish verified Accu-Chem's mathematical analysis utilizing those two calibrators, then
re-examined the veracity of Accu-Chem's results utilizing zero as an additional calibration
point. Dr. Wimbish testified that ideally, all four of the calibrators in Accu-Chem's test
would have been functional; however, changing the number of calibrators affects the test's
precision, but not its accuracy. In contrast, if all four calibrators fail, then the result is not
valid because there is no data with which to undertake any quantitative analysis. In short,
Dr. Wimbish testified that Accu-Chem's test rendered test results with reasonable scientific
certainty. Dr. Graham Richard Jones, a forensic toxicologist, testified on behalf of Alza.
According to Dr. Jones, the GC/MS is very powerful, but "it is basically a dumb instrument
unless you . . . calibrate it." Calibrators are blood samples containing known amounts of
fentanyl so that the machine can relate a signal to a concentration of fentanyl. According
to Dr. Jones, it is "critical" that the calibrators are working. Dr. Jones testified that multiple
calibrators are utilized in order to obtain multiple data points because the response across
a concentration range is not always consistent, and different calibrators are used to
establish the reliability of the test. According to Dr. Jones, the GC/MS machine may
"behave" differently at higher or lower concentrations; therefore, the calibrators should be
run at a range of concentrations within which the blood samples are expected to fall. In
his lab, Dr. Jones requires five out of six calibrators to be working to produce a quantitative
result for fentanyl. In other testimony, Dr. Jones stated that he might allow one or two
calibrators to fail; however, the calibrators must be dropped at the "extreme," or at a place
distant from where the actual quantitation was. In Dr. Jones's opinion, the Accu-Chem test
failed. Dr. Jones conceded that Dr. Wimbish disagreed with this analysis, and
acknowledged that Dr. Wimbish is a reputable forensic toxicologist and that Dr. Jones had
invited Wimbish to speak at a seminar on the subject of postmortem redistribution. Dr.
Jones further agreed that, in the instant case, the 10 quality control sample read close to
the target value, and further conceded that a limited sample of blood affects whether it is
appropriate to drop a calibrator. As lab director, he acknowledged that he has approved
a test where a distant calibrator was dropped because of a limited quantity of blood. Alza further contends that Accu-Chem's test was not properly performed because
Accu-Chem obtained its quality control material from the same source that it obtained its
calibrators. Dr. Wimbish addressed the "same source issue" with regard to Accu-Chem's
testing. Dr. Wimbish testified that it is generally recommended that a laboratory obtain its
quality control material from a source separate from the source used to obtain the
calibrators, but it is not required. Alza does not further explain or show how this adversely
affected the quality of the test or the test results itself. Finally, we note that Dr. Wimbish testified that Accu-Chem also tested Thompson's
blood for the metabolite of fentanyl and found less than 2 nanograms per milliliter. Based
on the amount of fentanyl in Thompson's system, Dr. Wimbish would have expected 8 or
10 nanograms per milliliter, and the discrepancy means that the fentanyl in Thompson's
system came in acutely, or over a short period of time. According to Dr. Wimbish, the "only
way" to explain the discrepancy between the amount of fentanyl and the amount of fentanyl
metabolites is that within a day's period, fentanyl came into Thompson's system in a "fairly
rapid" fashion in a "bolus amount," which is a "large amount . . . flooding into something at
the same -- at the same time." The trial court's gatekeeping function under Rule 702 does not supplant
cross-examination as "the traditional and appropriate means of attacking shaky but
admissible evidence." Gammill v. Jack Williams Chevrolet, 972 S.W.2d 713, 728 (Tex.
1998) (citing Daubert v. Merrell Dow Pharms., 509 U.S. 579, 596 (1993)); see Tex. R. Evid.
702. The reasons given by Alza as to why it thought the Accu-Chem test was unreliable
affect the weight and credibility of the test, but not its admissibility. See Mo.-Kan.Tex. R.R.
Co. v. May, 600 S.W.2d 755, 756 (Tex. 1980). Alza and its expert witnesses had the
opportunity to explore the consequences of these alleged defects in the testing procedures
and did so at great length. Once the trial court found that the scientific principles and procedures underlying the
test were generally accepted in the relevant scientific community, and the application of
this theory was reliable, the threshold determination of admissibility of the test was met,
and the judge was allowed, if not required, to allow the experts to render opinions on the
presence of fentanyl in Thompson's blood, leaving for the jury to determine the credibility
of those expert opinions. See Ledesma, 242 S.W.3d at 42. The issue of whether the
particular analysis of Thompson's blood by Accu-Chem was performed in the proper
manner, including whether sufficient quality control procedures were used to ensure the
accuracy of the test results, was a question of credibility and probative value for the jury
to determine, after hearing the testimony of all the experts. Alza contends that even if Accu-Chem's blood test was accurate, the level of
fentanyl in Thompson's blood was the result of postmortem redistribution rather than an
overdose. Dr. Wimbish testified that postmortem redistribution occurs after death when
body tissue and organs break down and drug concentrations in one organ may leak into,
or contaminate, other organs. In other words, for instance, if a drug is known to
concentrate in certain tissues, such as the lung, after death, the drug concentration may
migrate or leak from the lung tissue into the heart and cause an artificially elevated level
of the drug to be found in a blood sample taken from the heart. Dr. Wimbish testified that
all pathologists have been trained that the best place to collect a blood specimen to
prevent contamination is the femoral specimen, and it is scientifically reliable to form an
opinion about levels of fentanyl in someone's blood at the time of death based on levels
taken postmortem from a femoral draw. According to Dr. Wimbish, "[t]hat's been studied
on many occasions, and the femoral specimen is by far the best." Dr. Wimbish testified
that drugs that are highly protein bound and drugs that have high volumes of distribution
are those that are most susceptible to postmortem redistribution. If postmortem
redistribution were to occur, it would be significantly reduced by the refrigeration prior to
autopsy. Dr. Wimbish testified regarding an article entitled "Postmortem Redistribution of
Morphine and its Metabolites," published in the Journal of Forensic Sciences. Dr. Wimbish
testified that fentanyl and morphine are very similar compounds and are both considered
to be drugs that have reasonably high volumes of distribution, and accordingly, a study
regarding the postmortem redistribution of morphine would render results that could be
used to analyze whether fentanyl was susceptible to postmortem redistribution. According
to Dr. Wimbish, the morphine study analyzed blood samples taken from the femoral artery
and from heart blood to determine whether morphine is susceptible to postmortem
redistribution and determined that there was no significant difference in the results; thus,
postmortem redistribution of morphine from a human femoral draw was unlikely. This
study, like all others published in the Journal of Forensic Sciences, was subjected to peer
review, a critique by other scientists to evaluate and determine if the scientific integrity and
the protocol as designed in the study would allow a conclusion on a scientific basis. Dr. Wimbish testified that there are no peer reviewed, published articles concerning
postmortem redistribution of fentanyl from a human femoral draw; accordingly, theories
that postmortem fentanyl redistribution in humans regarding femoral and heart blood
specimens are speculation. Dr. Wimbish also testified about a study performed by Janssen Research
Foundation, which involved the application of Duragesic patches to six rabbits. This study
was not published or subject to peer review and was instead a "confidential" study by
Janssen, a private company and a subsidiary of the manufacturer of the Duragesic patch.
Janssen administered fentanyl to these rabbits through Duragesic transdermal patches and
collected blood specimens from the rabbits at different time intervals, both prior to and after
the death of the rabbits, to try to determine if postmortem redistribution occurred in the
rabbits. Although Janssen concluded that the study indicated that some postmortem
redistribution occurred with fentanyl, Dr. Wimbish disagreed. Dr. Wimbish testified that it was not scientifically reliable to derive an opinion about
the postmortem redistribution of fentanyl in human femoral blood from a test involving the
postmortem distribution of fentanyl in six four-and-a-half-pound rabbits and there was not
"any equivalence" to allow that correlation. Dr. Wimbish offered some scathing and
detailed criticisms of Janssen's study, which may be summarized as follows. This study
utilized a human patch on the body of a 4.5 pound rabbit. The "rabbit" sample was too
limited, and the concentrations in the blood and plasma ranged from therapeutic to toxic
in three of the six rabbits. The dosages of fentanyl were not administered on a milligram
per killigram basis, whereby the amount of drug administered to the rabbits could be
compared to the amount of drug typically prescribed to a human. The surface area of the
patch proportionally covered a much larger portion of the rabbits than it does humans.
Utilizing the study data, Dr. Wimbish testified that it shows no real difference in the initial
blood concentrations and the allegedly increased blood concentrations. While there are
some increases in concentrations, the Duragesic patches were left on the rabbits after their
deaths and the continued infusion of fentanyl could cause a rise in fentanyl levels
throughout the body after death. Dr. Wimbish stated that this process could not be
differentiated from what Alza's experts identified as postmortem redistribution. In short,
Dr. Wimbish testified that this study gives no support to the theory that postmortem
redistribution of fentanyl occurs in human femoral blood. Dr. Wimbish further testified about an article entitled "Duragesic Transdermal Patch:
Postmortem Tissue Distribution of Fentanyl in 25 Cases," published in the Journal of
Analytical Toxicology by Anderson and Muto. This peer-reviewed study analyzed twenty-five cases involving humans wearing fentanyl patches where death occurred. According
to Dr. Wimbish, this study did not show that there was any postmortem redistribution. The
authors referenced five cases indicating that the postmortem level of fentanyl in the
femoral blood was higher than in the central blood. However, according to Dr. Wimbish,
the paper was not designed to test whether postmortem redistribution occurred, and it does
not support a scientifically reliable theory that it occurs. On cross-examination, Dr.
Wimbish agreed that the study states that: "As with other drugs, the amount of
postmortem redistribution appears to increase as the Fentanyl concentration increases." Dr. Jones testified that in his opinion, the Accu-Chem test was not valid, and even
if the 11.4 level was accurate, that the blood level shown 18 hours after death did not
accurately reflect the level of fentanyl in Thompson's system at the time of death. Dr.
Jones testified that fentanyl is susceptible to postmortem redistribution, and the Anderson
and Muto study stands for the proposition that the amount of postmortem redistribution
appears to increase as the fentanyl concentration increases. According to Dr. Jones,
postmortem redistribution occurs most strongly around the lungs and heart and less so
peripherally in locations like the femoral artery. However, draws from the femoral artery
can still be affected unless it is a "closed" draw. According to Dr. Jones, the Anderson and
Muto study shows that postmortem redistribution occurs with fentanyl but "maybe in not
every case." He agreed that the scientific community had not reached a conclusion as to
whether postmortem redistribution artificially elevates fentanyl concentrations in human
femoral blood. Dr. Jones further testified that resuscitation efforts can increase the degree
of postmortem redistribution, and noted that testimony from nurse Kathryn Nu Bourn
indicated that CPR was performed on Thompson for approximately twenty minutes prior
to death. Dr. Jones testified that the "rabbit" study performed by Janssen indicates that
fentanyl concentrations increase after death, and "the most likely cause or contributor
would be postmortem redistribution." Dr. Jones first became aware of the Janssen rabbit
study when he was provided that study by Janssen's counsel. He conceded that it was
unpublished and was not peer-reviewed, and that it was performed "confidentially" by the
defendant that has been sued in this case. He acknowledged that the four-to-five pound
rabbits were subjected to more than twenty-two times the level of fentanyl that is used in
humans, and that the Duragesic patches remained on the rabbits after death and
continued to distribute fentanyl into the rabbits. Dr. Jones was retained by Janssen as an expert in at least sixteen cases. He
expects that in "certainly most cases" he will testify that the fentanyl levels are too high
because postmortem redistribution "could have occurred." He cannot say how much
Thompson's fentanyl blood level would have been affected by postmortem redistribution
to calculate what the fentanyl level would have been at the time of death. Based on the foregoing testimony, we cannot conclude that Alza's evidence
conclusively established that the level of fentanyl in Thompson's blood was the result of
postmortem redistribution. At most, the evidence presented in support of this theory was
speculative in nature. Alza's own expert, Dr. Jones, stated that postmortem redistribution
of fentanyl in humans "could have occurred," but "maybe in not every case," and he
conceded that the scientific community had not yet determined if postmortem redistribution
artificially elevates fentanyl concentrations in human femoral blood. The studies utilized
by Alza to support its theory were either not directed at determining whether postmortem
redistribution occurs with fentanyl, such as the Anderson and Muto study, or were
"confidential" studies, which were not peer reviewed, which were performed by Alza or its
own subsidiaries or employees. The study that Alza primarily relies on to support its theory
of postmortem redistribution relied solely on fentanyl testing on six rabbits. Animal studies,
standing alone, are generally held to be inadequate to prove causation in humans absent
other confirming epidemiological data. See, e.g., Havner, 953 S.W.2d at 729 (collecting
authorities). Accordingly, we reject Alza's contention that Thompson's fentanyl blood level
of 11.4 ng/ml was conclusively caused by postmortem redistribution. Alza contends that the evidence conclusively establishes that Thompson could not
have been wearing a 75 mcg/h patch from a recalled lot when she died. (22) First, Alza
argues that the Thompson family cannot prove that Michaelynn Thompson's Duragesic
patch was among the 2,500,000 patches recalled either during the initial recall or the
subsequent expanded recalls that Alza had to perform when it became apparent that its
initial recall was not nearly broad enough. However, proof that Thompson's Duragesic
patch was itself recalled is not an element of the Thompson family's claims and is,
therefore, immaterial. Under the Texas Rules of Evidence, "written notification by a manufacturer of any
defect in a product produced by such manufacturer to purchasers thereof is admissible
against the manufacturer on the issue of existence of the defect to the extent that it is
relevant." Tex. R. Evid. 407(b); Parsons, 85 S.W.3d at 331 n.2; Wright v. General Motors
Corp., 717 S.W.2d 153, 155 (Tex. App. Houston 1st Dist. 1986, no writ). While evidence
that a product fell within a recall is evidence of a defect, the converse is not true: evidence
that a manufacturer did not recall a product or evidence that the product at issue fell
outside the recall is no evidence that the product is not defective. If the Thompson family
had brought claims that focused on an allegation that Alza had negligently executed a
recall which it undertook to perform, (23) evidence that Michaelynn Thompson's Duragesic
patch was among the 2,500,000 patches recalled might possibly be of some relevance, but
Alza's argument is wholly irrelevant to the Thompson family's strict product liability
manufacturing defect claims (and also irrelevant to the family's general negligence claims,
which focused on Alza's careless quality assurance processes). Although the emergency room doctor removed and discarded Thompson's
Duragesic patch before it could be analyzed, it has been a well accepted principle of Texas
law for at least forty years that, in products liability cases, the claimant need not offer direct
evidence to prove a defect because such proof "usually can only be made by circumstantial
evidence." Turner v. Gen. Motors Corp., 584 S.W.2d 844, 848 (Tex. 1979) (citing Darryl
v. Ford Motor Co., 440 S.W.2d 630, 632 (Tex. 1969)); see Ridgway, 135 S.W.3d at 602
(Hecht, J., concurring) (stating that Texas law allows proof of product liability by
circumstantial evidence in certain cases); see also Shaun T. Mian Corp. v.
Hewlett-Packard Co., 237 S.W.3d 851, 858 (Tex. App.-Dallas 2007, pet. denied)
(providing a thorough examination of the principles underlying the use of circumstantial
evidence in product liability cases). Accordingly, when "the plaintiff has no evidence of a
specific design defect or manufacturing defect, he may offer evidence of the product's
malfunction as circumstantial proof of the defect." Sipes v. Gen. Motors Corp., 946 S.W.2d
143, 155 (Tex. App.-Texarkana 1997, writ denied). In this case, it is undisputed that Alza's Bodolay machine was the machine which
would have made Thompson's Duragesic patch as well as the many patches which had
documented defects in their seals. (24) This Bodolay machine was already scheduled for
replacement, but before it could be replaced, it had to undergo several modifications to try
to reduce the number of manufacturing defects it was producing. It is undisputed that
these modifications were not implemented until after the time period when Thompson's
Duragesic patch would have been manufactured on Alza's Bodolay machine. Moreover,
the documentary evidence and Alza's own witnesses confirmed that public reports of
leaking patches were higher in some lots of Duragesic patches that were not recalled than
in some of the lots that Alza recalled. Finally, Duragesic patches with defective seals are
a known cause of patients receiving an overdose of fentanyl, and as stated in Alza's own
second recall notice, it is undisputed that such fentanyl exposure can be fatal: Exposure to the leaked medication could result in . . . an increased
transdermal absorption of the opiate component fentanyl, leading to
potentially life-threatening complications. Based on the foregoing, and other evidence discussed herein, the jury had before it legally
and factually sufficient evidence from which it may have concluded that a manufacturing
defect in the Duragesic patch worn by Thompson caused her death. Accordingly, we
overrule Alza's first issue and those subissues therein. In its third issue, Alza contends that there was legally and factually insufficient
evidence to support the jury's causation findings. Alza also challenges the reliability of the
expert testimony supporting the causation finding. See Volkswagen of Am., Inc. v.
Ramirez, 159 S.W.3d 897, 903 (Tex. 2004)("A party may raise a properly preserved
complaint on appeal that scientific evidence is unreliable and thus no evidence to support
a judgment."). First, we address the reliability of the expert testimony, and then we will
separately address the sufficiency of the evidence. See Whirlpool Corp. v. Camacho, 298
S.W.3d 631, 639 (Tex. 2009); Volkswagen of Am., Inc., 159 S.W.3d at 904. Alza argues that the testimony offered by Thompson's experts, Dr. Robert C. Bux
and Dr. Sayed Feghali, was legally and factually insufficient and it varied materially from
the actual "undisputed" facts. Alza contends that, because of these deficiencies,
Thompson failed to produce expert testimony that any defect in a Duragesic patch worn
by Thompson was a producing or proximate cause of her death. Alza further argues that
the causation opinions of Thompson's experts are no evidence because they are based
solely on the Accu-Chem test. In this context, Alza argues that Thompson's expert opinion
testimony fails the Robinson test and is unreliable. See E.I. du Pont de Nemours & Co.
v. Robinson, 923 S.W.2d 549, 557 (Tex. 1995). A two-part test governs whether expert testimony is admissible: (1) the expert must
be qualified; and (2) the testimony must be relevant and based on a reliable foundation.
Helena Chem. Co. v. Wilkins, 47 S.W.3d 486, 499 (Tex. 2001). Whether the trial court
properly admitted expert testimony is subject to an abuse of discretion standard of review.
Id.; Robinson, 923 S.W.2d at 549, 558. The test for abuse of discretion is whether the trial
court acted without reference to any guiding rules or principles. Id.; Downer v. Aquamarine
Op., Inc., 701 S.W.2d 238, 241-42 (Tex. 1985). The test is not whether "in the opinion of
the reviewing court, the facts present an appropriate case for the trial court's action."
Robinson, 932 S.W.2d at 558. A reviewing court cannot conclude that a trial court abused
its discretion merely because it would have ruled differently. Id. The trial court has "broad"
discretion to determine the admissibility of expert evidence. Exxon Pipeline Co. v. Zwahr,
88 S.W.3d 623, 629 (Tex. 2002); Sanchez, 997 S.W.2d at 590; Robinson, 923 S.W.2d at
558. We must uphold the trial court's evidentiary ruling if there is any legitimate basis for
it. Owens-Corning Fiberglas Corp. v. Malone, 972 S.W.2d 35, 43 (Tex. 1998); Norstrud
v. Trinity Universal Ins. Co., 97 S.W.3d 749, 752 (Tex. App.-Fort Worth 2003, no pet.). For an expert's testimony to be admissible, it must possess a reliable foundation.
Cooper Tire & Rubber Co., 204 S.W.3d at 800. "Admission of expert testimony that does
not meet the reliability requirement is an abuse of discretion." Id. Expert testimony is
unreliable if it is based on unreliable data, or if the expert draws conclusions from his
underlying data "based on flawed methodology." Havner, 953 S.W.2d at 714. Expert
testimony is also unreliable if "there is simply too great an analytical gap between the data
and the opinion proffered." Gammill, 972 S.W.2d at 726 (quoting Gen. Elec. Co. v. Joiner,
522 U.S. 136, 146 (1997)). In E.I. du Pont de Nemours & Co. v. Robinson, the supreme court set out six factors
courts may consider in deciding whether expert testimony is reliable: 1. the extent to which the theory has been or can be tested; 2. the extent to which the technique relies on the subjective
interpretation of the expert; 3. whether the theory has been subjected to peer review and/or
publication; 4. the technique's potential rate of error; 5. whether the underlying theory or technique generally has been
accepted as valid by the relevant scientific community; and 6. the nonjudicial uses which have been made of the theory or
technique. 923 S.W.2d at 557. These Robinson factors are nonexclusive, Cooper Tire & Rubber Co.,
204 S.W.3d at 801, and "are not always useful in evaluating expert testimony." Id. at 802.
When the Robinson factors do not readily lend themselves to a review of the expert
testimony, "there must be some basis for the opinion offered to show its reliability."
Gammill, 972 S.W.2d at 726. An expert's "bare opinion" will not suffice and is unreliable
if "based solely upon his subjective interpretation of the facts." Volkswagen of Am., Inc.,
159 S.W.3d at 906. Dr. Bux, a physician and board certified forensic pathologist who practices as a
medical examiner and the coroner for El Paso County, Colorado, provided testimony for
the Thompson family. Based on his review and analysis of Thompson's medical records
from the day of her death, her autopsy report and microscopic pathology slides, and the
Accu-Chem data, Dr. Bux testified that Thompson died from a fentanyl overdose and not
heart disease. Thompson presented to the emergency room with symptoms of nausea,
vomiting, and back pain, and later complained of dizziness, but she was alert and oriented.
She was on an EKG monitor during her transportation to the hospital by the emergency
medical personnel, and also while she was in the emergency room, and "those monitors
do not reveal any kind of cardiac abnormality or rhythm disturbance" but instead show "her
regular normal heartbeat." Thompson's medical records from the day of her death do not
show that she reported any symptoms that would suggest that she was experiencing a
cardiology-related problem, such as chest pain or shortness of breath. Dr. Bux further testified that a "cardiac panel" test was performed on Thompson's
blood. This panel included an array of tests, which, inter alia, checked for high levels of
certain enzymes indicating damage to muscles, including cardiac muscle, and troponin,
which is specific for cardiac muscle. The results of the cardiac panel were "all either
negative or within normal limits, which tells me that at the time she came in, it did not
appear from an enzyme standpoint that she had anything going on with her heart." Finally,
examining the autopsy data, Dr. Bux testified that, microscopically, Thompson's heart
muscle looked normal except for the hypertrophy. He concluded there was no forensic
evidence in the records on the day that Thompson died that supports a determination that
Thompson died from sudden cardiac arrest and that sudden cardiac arrest was not a
plausible cause of Thompson's death. According to Dr. Bux, a large heart is not an actual
cause of death. Dr. Bux considered the other drugs that were administered to Thompson at the
hospital, including Antivert, Zofran, and Dilaudid, and excluded them as factors in
Thompson's demise based on the dosages given, the methods of infusion, and the timing
of their administration. He found it significant that the emergency room physician removed
the Duragesic patch from Thompson during his efforts to resuscitate her and also
administered Narcan, which is an antidote for a narcotic overdose. Dr. Bux testified that Accu-Chem's toxicological report showed a quantitation of
fentanyl at 11.4 ng/ml, which is a "lethal" level. He further testified that respiratory
depression due to a drug overdose occurs in the last minutes or hour before death, that
it is very difficult to "pick up," and that it is completely foreseeable for Thompson to not
show signs of respiratory depression at 3:15 p.m., but die from it thirty-five minutes later.
Dr. Sayed Feghali, a cardiologist who works at the Texas Heart Institute and
teaches at Baylor School of Medicine, also testified on behalf of the Thompsons regarding
whether or not Thompson could have died from cardiac problems. (25) He is board certified
in cardiology and interventional cardiology. This was his first time to provide expert
testimony in a lawsuit. According to Dr. Feghali, fentanyl has a narrow therapeutic margin. He has used
it intravenously in thousands of patients. After reviewing all of Thompson's medical
records, including EKGs from a sleep study that Thompson had participated in, the EKG
taken by emergency workers on the day Thompson died, and the EKGs taken during
attempts to resuscitate her, Dr. Feghali testified that Thompson died from an overdose of
fentanyl, and Thompson's morbid obesity and sleep apnea played a minor part, if any, in
her death. Dr. Feghali testified that other medications that she was given could not have
caused respiratory depression because she was given appropriate doses. Dr. Feghali concluded that Thompson could not have died from a cardiac event
because her autopsy showed no damaged heart tissue and the pathology report
referenced no damage to Thompson's heart muscle or any blockage of her arteries.
According to Dr. Feghali, a sudden malignant arrhythmia does not come from nowhere and
there should be a cause for it somewhere in the heart, like heart damage, heart scarring,
a blockage, or a viral infection. In other words, the heart muscle will show prior damage
of some kind. Dr. Feghali further based his conclusion that Thompson did not die from a
sudden cardiac event because electric discrepancies did not appear on the EKGs
performed either while en route to the emergency room, or saliently, while the emergency
team was trying to resuscitate her. Dr. Feghali testified that while Thompson's heart was
large, it was normally sized for her body. On appeal, Alza does not challenge Dr. Bux or Dr. Feghali's credentials. Their
testimony analytically linked the data they observed to their theory of how Thompson died.
They based their testimony on witness testimony, medical records, autopsy records, blood
samples taken at the autopsy, the blood testing, and Alza's own literature and drug
warnings. The testing was based on generally acceptable scientific principles. The jury
was free to examine this evidence. The experts' testimony does not present a case where
"there is simply too great an analytical gap between the data and the opinion proffered,"
Gammill, 972 S.W.2d at 726, or where the expert's testimony amounted to nothing more
than a recitation of his credentials and a subjective opinion. See Cooper Tire, 204 S.W.3d
at 801. The Thompson family met its burden to prove Dr. Bux and Dr. Feghali's testimony
was relevant and reliable. See Whirlpool Corp., 298 S.W.3d at 638. We conclude that
Alza's complaints about the experts' testimony go to its weight, not its admissibility. The
trial court therefore did not abuse its discretion in admitting the testimony of either Dr. Bux
or Dr. Feghali. We next examine the legal and factual sufficiency of the evidence of causation
supporting the jury's finding. See id. In addition to the evidence and testimony previously
recounted herein, we will examine the testimony of Alza's retained expert witnesses: Dr.
Vincent DiMaio, the medical examiner in Bexar County; Dr. Joseph Varon, who specializes
in internal medicine, intensive care, pulmonary and chest diseases, sleep disorders,
geriatrics, and emergency medicine; and Dr. Aaron Kenneth Calodney, a medical doctor
and pain management specialist. Dr. DiMaio testified that Thompson died from a fatal cardiac arrhythmia, an irregular
beating of the heart, produced by the massive enlargement of her heart. According to Dr.
DiMaio, it was essentially a "coincidence" that Thompson went to the hospital and died of
an arrhythmia. Dr. DiMaio testified that thirteen percent of people who die suddenly and
unexpectedly have a large heart as the cause of death. Yet he conceded before the jury
that one cannot perform a test for cardiomyopathy and he is not a cardiologist. While Dr.
DiMaio agreed that Thompson's medical records and testimony evidence sleepiness, and
sleepiness is a sign of fentanyl toxicity, he stated that if Thompson had received a fatal
dose of fentanyl, she would have become sleepy and then gone into a coma for two to
three hours prior to death, in contrast to the evidence herein which indicates that
Thompson was found dead thirty to thirty-five minutes after being seen by a nurse. Dr.
DiMaio testified that the only way a fentanyl overdose kills immediately is if it is infused
intravenously. Finally, Dr. DiMaio testified that, in retrospect, the medical examiner for
Harris County should have tested Thompson for fentanyl. Dr. Varon testified in counterpoint to Dr. Feghali. While Dr. Varon conceded that
he is not board-certified in cardiology, he claimed that he is the "number one guy" on
sudden cardiac death. He works at the medical center in Houston and teaches at the
University of Texas Health Science Center and other places. According to Dr. Varon, signs
to look for in diagnosing a fentanyl or narcotic overdose include: constricted pupils,
sweating, drowsiness or lethargy, and the cessation of breathing. Dr. Varon testified that
these symptoms appear fast with intravenous transfusions of fentanyl or more slowly when
the drug is introduced orally or through a skin patch. With a fentanyl overdose, he would
have expected Thompson to be lethargic, sleepy, drowsy, and have depressed
respirations, yet he did gave sworn testimony that he did not recall any of these symptoms
in her medical records which expressly document some of these symptoms. According to Dr. Varon, it takes sixteen hours to achieve the full blood level effects
of a fentanyl patch. However, when questioned about proper fentanyl dosages, he testified
that he does not follow the therapeutic ranges on narcotics in the PDR, and in response
to a question regarding whether his dosages regarding fentanyl were "catch as catch can,"
replied, "exactly right." Dr. Varon testified that he is familiar with sudden cardiac arrest, and that 500 to
1,000 cases of sudden cardiac arrest occur every day in the United States and that one
episode of cardiac arrhythmia can cause death. The underlying arrhythmia can be caused
by the heart not getting enough oxygen or potassium. According to Dr. Varon, there are
several risk factors for a sudden cardiac arrhythmia, including blockages in the heart, sleep
apnea, medications, problems with electrolytes, trauma, or an enlarged heart. Dr. Varon examined the autopsy report, which stated that Thompson was six feet
tall and weighed 367 pounds, and testified that obesity also carries a higher risk for health
problems like apnea and sudden cardiac death. Thompson's heart weighed 600 grams
and had enlarged chambers, whereas a normal heart weighs between 350 to 500 grams.
According to Dr. Varon, the more dilated the heart is, the higher the chances of cardiac
arrhythmia. Dr. Varon testified that in his opinion, Thompson did not have a healthy heart,
and he opined that she expired from a sudden cardiac death, which was the result of
arrhythmia or a defibrillate. He also testified that it was just a coincidence that Thompson
was in the hospital when she died of arrhythmia. In this regard, he stated that Thompson
had participated in a sleep study which showed she had sleep apnea, although he
conceded that the records from that study showed that she had 144 instances of apnea
without any corresponding arrhythmia. Dr. Varon testified that Thompson's medical
records from 2001 and 2002 show potential heart problems in two instances where she
complained of chest pains and tightness, difficulty breathing, nausea, and general
weakness, although the records fail to show any such diagnosis. According to Dr. Varon, an individual can experience a regular heart rate which
progresses to full arrhythmia in minutes. However, Dr. Varon admitted that the EKGs
which were performed on Thompson on the date of her death showed no abnormalities
with regard to Thompson's heart valves, and her autopsy showed no atherosclerosis or
other heart damage. He further agreed that big people can have big, healthy hearts. Dr. Aaron Kenneth Calodney, a medical doctor and pain management specialist,
provided further testimony on behalf of Alza. Dr. Calodney has written thousands of
prescriptions for Duragesic patches. He testified that Duragesic patches fall within a "class
II" system of narcotics, and a class II narcotic is something that is "[d]efinitely" dangerous
if misused. Dr. Calodney testified that he would not start an opioid naive patient on a 50
milligram patch because it could potentially be fatal. He testified that fentanyl is an opioid
like morphine, but is approximately 100 times more potent than morphine. Dr. Calodney testified that the PDR is a common reference for professionals to use
to ascertain the particular qualities and properties of drugs, and further testified regarding
the PDR's section on the Duragesic patch. The PDR states that "since the peak fentanyl
levels occur between 24 and 72 hours of treatment, prescribers should be aware that
serious or life-threatening hypoventilation may occur, even in opiate tolerant patients during
the initial application period." It further states that "[o]verestimating the Duragesic dose
when converting patients from another opiate medication can result in fatal overdose with
the first dose." The 2005 PDR states that using damaged or cut patches can lead to the
rapid release of the contents of the patch and absorption of a potentially fatal dose of
fentanyl. Dr. Calodney has never seen a damaged patch, but he agreed that if the patch was
damaged and the drug were to escape, then that could lead to a rapid release of fentanyl.
He refused to testify regarding whether a rapid release could lead to a fatal overdose, "not
knowing whether the stuff inside the patch can actually be absorbed through the skin"
without the patch system itself. According to Dr. Calodney, Thompson was an appropriate candidate for use of a
Duragesic patch. Dr. Calodney stated that complications from an overdose of fentanyl
include itching, nausea and vomiting, difficulty urinating, and respiratory depression; he
reviewed Thompson's medical records and nevertheless stated that he saw no signs in her
medical records that indicated the appearance of fentanyl toxicity. He further testified that
if someone is receiving too much fentanyl, a change in their level of consciousness is
observable over a period of time; the patient will begin to get drowsy and develop changes
in respiration after that. According to Dr. Calodney, these changes happen slowly and
gradually as the medication is infused, presumably through a rate-controlled membrane. We note at this juncture that, in contrast, Dr. Wimbish testified that the respiratory center
adapts to drug tolerance and a patient may not experience respiratory depression until the
threshold for tolerance is exceeded, and it may be close to the fatal event in a tolerant
individual. In other words, respiratory depression is not necessarily observable in a tolerant
individual. According to Dr. Wimbish, it could be "part of the terminal event rather than a
harbinger of something to come." To a significant extent, deciding whether Thompson died from an enlarged heart or
a sudden arrhythmia, as alleged by Alza, or an overdose of fentanyl, as alleged by the
Thompson family, turned on the jury's examination of the evidence and its assessment of
the credibility of the various witnesses. We note that the jury was, of course, entitled to
weigh the credibility of these witnesses. Ledesma, 242 S.W.3d at 41 (holding that the jury
was entitled to take into account that an expert had worked for the manufacturer and had
never found a defect in a product). (26) Causation may be established by circumstantial evidence. Ridgway, 135 S.W.3d
at 601; Ford Motor Co. v. Cammack, 999 S.W.2d 1, 7 (Tex. App.-Houston [14th Dist.]
1998, pet. denied); J.K. and Susie L. Wadley Research Inst. & Blood Bank v. Beeson, 835
S.W.2d 689, 698 (Tex. App.-Dallas 1992, writ denied). This principle of Texas law has
been well recognized for decades. Gladewater v. Pike, 727 S.W.2d 514, 518 (Tex. 1987)
("Circumstantial evidence and inferences therefrom are a sufficient basis for a finding of
causation."); Farley v. M M Cattle Co., 529 S.W.2d 751, 755 (Tex. 1975) ("Proximate
cause, like any other ultimate fact, may be established by circumstantial evidence."); Lynch
v. Ricketts, 158 Tex. 487, 314 S.W.2d 273, 275-76 (Tex. 1958) ("It is well settled . . . that
negligence and causation, like any other ultimate fact, may be established by
circumstantial as well as direct evidence."); Peveto v. Smith, 134 Tex. 308, 133 S.W.2d
572, 576 (1939) (relying on circumstantial evidence to prove negligence and causation);
see also Tex. Elec. Coop. v. Dillard, 171 S.W.3d 201, 206 (Tex. App.-Tyler 2005, no pet.)
("Cause in fact . . . may be proven by circumstantial evidence . . . ."). The question of
causation is a fact question for the jury, and the jury has broad latitude to infer causation
from the circumstances surrounding an accident, especially when it is not possible to
produce direct proof of causation. J.K. and Susie L. Wadley, 835 S.W.2d at 698; Farley,
529 S.W.2d at 756. The plaintiff need not exclude all possibility that the accident occurred
other than as he alleges. Farley, 529 S.W.2d at 756. Rather, the plaintiff is only required
to convince the jury by a preponderance of the evidence that the accident occurred as
alleged. Renfro Drug Co. v. Lewis, 235 S.W.2d 609, 621 (1950); Gulf States Util. Co. v.
Dryden, 735 S.W.2d 263, 267 (Tex. App.- Beaumont 1987, no writ). In the specific context
of a strict product liability case, the claimants can support their theory of how the incident
occurred with circumstantial evidence. See, e.g., Ridgway, 135 S.W.3d at 601; Gen.
Motors Corp. v. Sanchez, 997 S.W.2d 584, 587 (Tex. 1999); Kindred v. Con/Chem, Inc.,
650 S.W.2d 61, 63 (Tex.1983). As stated previously, under the Texas Supreme Court's articulation of the standard
of review in City of Keller, in reviewing the legal sufficiency of the evidence, we view the
evidence in the light most favorable to the jury's findings, giving full credit to all favorable
evidence if any reasonable person could, and disregarding contrary evidence unless all
reasonable persons could not. City of Keller, 168 S.W.3d at 807. The evidence supporting
the jury's finding on causation was not completely lacking, nor was it beyond the court's
authority to admit it, nor was it a mere scintilla, nor was it conclusively disproved by
contrary evidence. See Uniroyal Goodrich Tire Co. v. Martinez, 977 S.W.2d 328, 334 (Tex.
1998); see also Whirlpool Corp., 298 S.W.3d at 638. Moreover, considering and weighing
all of the evidence in the case, the verdict was not so against the great weight and
preponderance of the evidence as to be manifestly unjust, shock the conscience, or clearly
demonstrate bias. See Golden Eagle Archery, Inc. v. Jackson, 116 S.W.3d 757, 761-62
(Tex. 2003). Based on the foregoing, we overrule Alza's third issue. In its second issue, Alza contends that there is legally insufficient evidence to
support the submission or jury finding that Alza was negligent because there was no expert
evidence on the standard of care. As an initial matter, we note that a party may pursue claims and seek damages on
alternative theories of liability. Tex. R. Civ. P. 48; Waite Hill Servs., Inc. v. World Class
Metal Works, Inc., 959 S.W.2d 182, 184 (Tex. 1998). When we have held that the
evidence is legally and factually sufficient to support the trial court's judgment on one of
multiple alternative causes of action, "it is unnecessary to address" the issues that concern
alternative theories that might have also served as a basis to affirm the judgment. Adams
v. H & H Meat Prods., Inc., 41 S.W.3d 762, 771 (Tex. App.-Corpus Christi 2001, no pet.)
(citing Tex. R. App. P. 47.1); see Nw. Mortgage, Inc. v. Salinas, 999 S.W.2d 846, 865 (Tex.
App.-Corpus Christi 1999, pet. denied) (stating that when "the trial court's judgment can
be supported on the basis of" one of multiple theories of recovery, "we need not address
these remaining points" which address alternative theories of recovery). Accordingly, we
need not address Alza's second issue, which challenges the legal sufficiency of the
evidence supporting the jury's finding of negligence as an alternative basis for affirming the
judgment. See Tex. R. App. P. 47.1. Nevertheless, even if we were to find it necessary to address this issue, we note that
the jury was instructed that: "Negligence" means failure to use ordinary care, that is, failing to do that
which a pharmaceutical manufacturer of ordinary prudence would have done
under the same or similar circumstances or doing that which a
pharmaceutical manufacturer of ordinary prudence would not have done
under the same or similar circumstances. Where an issue involves only general knowledge and experience rather than expertise, it
is a matter entrusted to the jury to decide. See GTE Sw., Inc. v. Bruce, 998 S.W.2d 605,
620 (Tex. 1999). In contrast, "[e]xpert testimony is necessary when the alleged negligence
is of such a nature as not to be within the experience of the layman." FFE Transp. Servs.
v. Fulgham, 154 S.W.3d 84, 91 (Tex. 2004) (quoting Roark v. Allen, 633 S.W.2d 804, 809
(Tex. 1982) (holding that diagnosis of skull fractures is not within the experience of the
layman)); see also Turbines, Inc. v. Dardis, 1 S.W.3d 726, 738 (Tex. App.-Amarillo 1999,
pet. denied) (holding that inspection and repair of an aircraft engine are not within the
experience of the layman); Hager v. Romines, 913 S.W.2d 733, 734-35 (Tex. App.-Fort
Worth 1995, no writ) (holding that operation of an aircraft and aerial application of herbicide
are not within the experience of the layman). In determining whether expert testimony is
necessary to establish negligence, Texas courts have considered whether the conduct at
issue involves the use of specialized equipment and techniques unfamiliar to the ordinary
person. Fulgham, 154 S.W.3d at 91. In light of the testimony and admissions by Alza's employees and representatives
regarding its manufacturing and quality inspection process for the Duragesic patch, the
determination regarding whether Alza breached the standard of care applicable to a
reasonable pharmaceutical manufacturer does not entail specialized knowledge or skill
unique to a scientific discipline and beyond the knowledge and experience of the average
jury. See, e.g, Worsham v. A.H. Robins Co., 734 F.2d 676, 685 (11th Cir. Fla. 1984)
(affirming a judgment against the manufacturer of the Dalkon Shield IUD on claims of
negligence despite the absence of expert testimony on the duty of care); Palmer v. A.H.
Robins Co., 684 P.2d 187, 209-10 (Colo. 1984) (same); cf. McNeil Pharm. v. Hawkins, 686
A.2d 567, 585 (D.C. 1996) (requiring competent expert testimony regarding the violation
of the statutes and regulations submitted to the jury in warnings case). On the contrary,
resolution of the negligence issue turns on an assessment of evidence relating to Alza's
decision to manufacture the Duragesic patch on a machine known to produce defective
patches and to test its products with a method known to be insufficient to ascertain the
absence of defect. The ultimate determination of whether Alza's conduct comported with
that degree of care which a reasonably prudent drug manufacturer would use under the
same or similar circumstances was well within the experience of persons of ordinary
intelligence. Thus, expert opinion evidence was not essential to the Thompson family's
claim in negligence. (27)
However, even if we were to determine otherwise, and conclude that expert
testimony was necessary, this requirement would have been met by the testimony
proffered by Alza's own representatives and employees. While the Thompson family might
not have presented a retained expert on pharmaceutical manufacturing to establish their
claim that Alza was negligent in its manufacturing process, Alza's own representatives and
employees, including a senior technician, the company representative, and its quality
assurance specialist, offered expert opinion testimony regarding Alza's acts and omissions
with regard to the Duragesic patch, from which credible and expert evidence the jury may
have concluded that Alza was negligent. These witnesses testified that Alza released
leaking patches to the public and only learned that its patches were defective because of
consumer complaints; that Alza wrongly determined that the 2001 fold-over incidents were
isolated events and accordingly, did not take corrective action that would have prevented
distribution of the 2004 recalled lots; that Alza's inspection plan for the product lots was
deficient; that Alza did not follow its own deficient plan and allowed distribution of trays of
patches that it knew contained defects; that Alza continued to release patches after both
learning of the leaks and determining the manufacturing flaw that resulted in the leaks; and
that Alza continued to implement a visual inspection process despite knowing that it was
ineffective to discover product defects. Ultimately, in the instant case, the primary facts were accurately presented to the
jury, and the jurors were as capable of understanding and drawing correct conclusions
from the facts as an expert witness. While there are cases that would fail without expert
testimony because the technical and scientific aspects of the case would result in a jury's
inability to comprehend the issues, Alza has not shown that the standard of care issue is,
by definition, a matter beyond the jurors or that this is "one of the rare causes of action in
which the law predicates recovery upon expert testimony." Salem v. U.S. Lines Co., 370
U.S. 31, 35 (1962). For the foregoing reasons, we overrule Alza's second issue. Alza's fourth and fifth issues attack the trial court's admission of evidence regarding
Accu-Chem's test and its recall of the Duragesic 75 mcg patch. We review the admission or exclusion of evidence under an abuse of discretion
standard. See In re J.P.B., 180 S.W.3d 570, 575 (Tex. 2005). Thus, evidentiary rulings
admitting or excluding evidence are committed to the trial court's "sound" discretion. See
Tex. Dep't of Transp. v. Able, 35 S.W.3d 608, 617 (Tex. 2000); City of Brownsville v.
Alvarado, 897 S.W.2d 750, 753 (Tex. 1995); Gee v. Liberty Mut. Fire Ins. Co., 765 S.W.2d
394, 396 (Tex. 1989). A trial court abuses its discretion if it acts without reference to any
guiding rules and principles or if the act complained of is arbitrary and unreasonable. City
of Brownsville, 897 S.W.2d at 754; see also Carpenter v. Cimarron Hydrocarbons Corp.,
98 S.W.3d 682, 687 (Tex. 2002). We must uphold a trial court's evidentiary ruling if there
is any legitimate basis in the record to support it. Owens-Corning Fiberglas Corp., 972
S.W.2d at 43. For the exclusion of evidence to constitute reversible error, the complaining party
must show that (1) the trial court committed error, and (2) the error probably caused the
rendition of an improper judgment. State v. Cent. Expressway Sign Assocs., No. 08-061,
53 Tex. Sup. Ct. J. 134, 2009 Tex. LEXIS 967, at **5-6 (Tex. Nov. 30, 2009); McCraw v.
Maris, 828 S.W.2d 756, 757 (Tex. 1992); see Tex. R. App. P. 44.1; Nissan Motor Co. v.
Armstrong, 145 S.W.3d 131, 144 (Tex. 2004). A person seeking to reverse a judgment
based on evidentiary error need not prove that "but for" the error a different judgment would
necessarily have been rendered, but only that the error probably resulted in an improper
judgment. State, 2009 Tex. LEXIS 967, at *6; City of Brownsville, 897 S.W.2d at 754;
McCraw, 828 S.W.2d at 758. A successful challenge to evidentiary rulings usually requires
the complaining party to show that the judgment turns on the particular evidence excluded
or admitted. Bay Area Healthcare Group, Ltd. v. McShane, 239 S.W.3d 231, 234 (Tex.
2007); Nissan Motor Co., 145 S.W.3d at 144. In making this determination, we review the
entire record. Reliance Steel & Aluminum Co. v. Sevcik, 267 S.W.3d 867, 871 (Tex. 2008);
Nissan Motor Co., 145 S.W.3d at 144. The supreme court has recognized the impossibility of prescribing a specific test
for harmless error review and has characterized the standard as "more a matter of
judgment than precise measurement." Reliance Steel, 267 S.W.3d at 871 (quoting Nissan
Motor Co., 145 S.W.3d at 144). Accordingly, as stated by Justice O'Neill, that
determination is entrusted "to the sound discretion of the reviewing court." State, 2009 Tex.
LEXIS 967, at *5 (citing McCraw, 828 S.W.2d at 757-58; Lorusso v. Members Mut. Ins.
Co., 603 S.W.2d 818, 821 (Tex. 1980)). We consider several factors in conducting a harmless error review for the erroneous
admission of evidence: the state of the evidence, the strength and weakness of the case,
and the verdict. Reliance Steel, 267 S.W.3d at 871. Stated otherwise, we consider the
effect of the admission of the evidence; the role the evidence played in the context of trial;
efforts by counsel to emphasize the erroneous evidence; and whether the admission of the
improper evidence was calculated or inadvertent. See id. at 871-75. Thus, "the exclusion
or admission is likely harmless if the evidence was cumulative, or the rest of the evidence
was so one-sided that the error likely made no difference in the judgment." State, 2009
Tex. LEXIS 967, at *6; see Reliance Steel, 267 S.W.3d at 873. But if erroneously admitted
or excluded evidence was "crucial" to a "key issue," the error is likely to be found harmful.
State, 2009 Tex. LEXIS 967, at *6. We further consider whether there was contrary
evidence that the improperly admitted evidence was calculated to overcome. Id;
Armstrong, 145 S.W.3d at 144. In its fourth issue, Alza contends that the trial court committed harmful error resulting
in an improper judgment by admitting the results of the Accu-Chem postmortem blood test
that was improperly performed and by permitting Thompson's experts to opine on the
cause of death based solely on unreliable test results. We have previously addressed this
matter with regard to Alza's other issues and need not further address it herein. See Tex.
R. App. P. 47.1. This issue is overruled. In its fifth issue, Alza contends that the trial court's admission of evidence that Alza
recalled several lots of Duragesic patches constituted an abuse of discretion that was
reasonably calculated to and probably caused the rendition of an improper judgment. In
several sub-issues, Alza argues that: (1) there is no evidence independent of the recall
that Thompson was wearing a patch with the same defect; (2) the evidence of recall does
not tend to make a fact that is of consequence to the suit any more or less probable; (3)
the evidence of recall was outweighed by prejudice and caused a confusion of the issues;
and (4) the admission of the recall evidence was reasonably calculated to and probably
caused the rendition of an improper judgment. Thompson counters that this evidence was
both admissible and relevant. On February 17, 2004, Alza recalled one lot of 75 mcg Duragesic patches through
an "Urgent Class 1 Drug Recall Notification." This recall notice stated that a "potential seal
breach on one edge may allow drug to leak from the patch." The notice further stated that: Exposure to the Duragesic hydrogel contents could result in an increased
absorption of the opioid component, fentanyl, leading to increased drug
effect, including nausea, sedation, drowsiness, or potentially life threatening
complications. On April 7, 2004, Alza recalled four additional lots of 75 mcg Duragesic patches in an
"Urgent Expanded Drug Recall Notification." This recall warned that: Exposure to the leaked medication could result in inadvertent ingestion or an
increased transdermal absorption of the opiate component, fentanyl, leading
to potentially life-threatening complications. Texas Rule of Evidence 407 provides: A written notification by a manufacturer of any defect in a product produced
by such manufacturer to purchasers thereof is admissible against the
manufacturer on the issue of existence of the defect to the extent it is
relevant. Tex. R. Evid. 407(b). This Court addressed the admissibility of recall evidence in Ford
Motor Co. v. Durrill, 714 S.W.2d 329 (Tex. App.-Corpus Christi 1986), judgm't vacated by
agreement, 754 S.W.2d 646 (Tex. 1988): In appellant's twentieth point of error, Ford urges that the trial court erred in
admitting the Pinto recall letter and testimony relating to the Pinto recall. The
Durrills argue that the evidence was relevant to show Ford's conscious
disregard for the public safety. Rule 407(b) allows a written ratification by a
manufacturer of a product defect to be admitted against the manufacturer on
the issue of existence of the defect to the extent that it is relevant. The fact
that recall letters are admissible is settled by the existence of Tex. R. Evid.
407(b). The circumstances for admission in this case are somewhat unusual
in that the letter was admitted to show both Ford's failure to warn of an
alleged similar defect in the Mustang II and to show that a defect existed in
the Mustang II by showing its similarities to the recalled Pinto. Dr. Robert Brenner testified that the Ford Motor Company should have sent
a recall letter like the "Pinto" letter to Mustang owners as well. He testified
that the safety related defect which precipitated the Pinto -Bobcat recall was
very similar, if not identical to the safety related defect that Ford knew existed
in the Pinto-Bobcat and Mustang. The evidence of the Pinto recall was relevant both to the defect issue and
the failure to warn. It was incumbent upon the Durrills to show by
independent evidence that the Mustang II was defective at the time of injury
in order for the letter to be admissible to show the existence of a defect. See
Blakely, Article IV: Relevancy and Its Limits, 20 Hous. L. Rev. 151, 220
(1983). The Durrills established, by independent evidence, that the Mustang
II was defective in the same manner as the Pinto. Additionally, we find that
the evidence was admissible to illustrate Ford's failure to warn. See id. at 340. Rule 407(b) clearly states that the evidence of the recalls is admissible in evidence
on the issue of the existence of the defect to the extent that it is relevant. Specifically, the
issue of recall is, in the words of Texas Rule of Evidence 407, "admissible against [Alza]
on the issue of existence of the defect to the extent it is relevant." The defect referenced
in the recall is the same defect as alleged in the instant case, that is, an alleged breach in
the seal of the patch causing an uncontrolled release of fentanyl, and thus, the recall is
relevant. Alza contends that there is no evidence independent of the recall that Thompson
was wearing a patch with the same defect as referenced in the recall. Assuming without
deciding that Thompson was required to show some independent proof that the particular
product in question suffered from the same defect, the record contained such evidence.
Thompson was prescribed the same patch that was the subject of Alza's recall, and her
patch was manufactured during the same period of time and on the same machine as the
recalled patches. Prior to and upon admission to the hospital, she displayed symptoms
of overdose including nausea, vomiting, sleepiness, and dizziness. Thompson's blood test
showed an overdose of fentanyl. Accordingly, there is independent evidence of the alleged
defect in the involved product. Alza contends that the evidence of recall does not tend to make a fact that is of
consequence to the suit any more or less probable. However, recall letters are admissible
as some evidence that the defect existed when the product left the manufacturer. This is
an element that Thompson was required to prove. Ridgeway, 135 S.W.3d at 600 (holding
that the plaintiff must prove that the product was defective when it left the hands of the
manufacturer and that the defect was a producing cause of the plaintiff's injuries). Finally, Alza contends that the evidence of recall was outweighed by prejudice and
caused a confusion of the issues, and the admission of the recall evidence was reasonably
calculated to and probably caused the rendition of an improper judgment. Because
Thompson was required to prove that the defect existed not only at the time of her death
but also at the time when the Duragesic patch left Alza's hands, the relevancy of the recall
letters outweighs any possible prejudice in their admission into evidence. In this regard,
we note that "testimony is not inadmissible on the sole ground that it is 'prejudicial' because
in our adversarial system, much of a proponent's evidence is legitimately intended to
wound the opponent." Bay Area Healthcare Group, 239 S.W.3d at 234. A jury is entitled
to know that the existence of the defect in a particular kind of Duragesic patch was likely
and that such likelihood was greater as to the 75 mcg patches as to other patches
manufactured by Alza. Accordingly, the prejudicial impact of the recall letters is
outweighed by its relevance on the issue whether the product was defective when it left the
hands of the manufacturer. We note that the recall does not by itself make a prima facie
case, shift the burden of proof, or prove that the defect existed at the time of the incident.
See, e.g., Fields v. Volkswagen of Am., Inc., 555 P.2d 48, 58 (Okla. 1976), We conclude that, under the circumstances of this case, the recalls were properly
admissible in evidence for the purpose of determining whether the defect identified by
Thompson's expert witnesses existed when Alza put the patch into Thompson's hands.
We note that to sustain a verdict for a plaintiff based upon an existing defect, such recall
evidence, standing alone, would be insufficient. Here, however, we have the credible
testimony of Thompson's experts, albeit contradicted by that of Alza's expert witnesses,
which justified the jury's conclusion that the Duragesic patch worn by Thompson was
defective and the defect caused her death. Accordingly, we hold that, in the circumstances
of this case, the recall evidence was properly admissible in evidence for the limited
purpose of determining whether the defect identified by plaintiff's expert witnesses arose
while the product was in Alza's control. We overrule Alza's fifth issue. Having overruled each of Alza's issues, we affirm the judgment of the trial court.
_____________________ Linda R. Yañez Justice Delivered and filed the 1st day of April, 2010. 1. 2. "While strict liability focuses on the condition of the product, '[n]egligence looks at the acts of the
manufacturer and determines if it exercised ordinary care in design and production.'" Am. Tobacco Co., Inc.
v. Grinnell, 951 S.W.2d 420, 437 (Tex. 1997) (quoting Caterpillar, Inc. v. Shears, 911 S.W.2d 379, 384 (Tex.
1995)).
3. Alza's brief describes fentanyl as "a powerful narcotic painkiller" administered by a "Duragesic patch
that releases doses of the medicine into the patient's bloodstream via the skin."
4. We note that the trial court was not the first court to address punitive damage claims against Alza
based on allegedly defective Duragesic fentanyl patches. See, e.g., Lake-Allen v. Johnson & Johnson, L.P.,
No. 2:08CV00930DAK, 2009 U.S. Dist. LEXIS 64860, at *9-10 (D. Utah July 27, 2009) (denying defendants'
motion for summary judgment on punitive damage claims).
5. Plaintiffs below and appellees herein include Doris Thompson, individually, and Townsi Foulkrod as
legal representative of Kenzey Thompson, minor, individually and on behalf of the estate of Michaelynn
Thompson.
6. The evidence showing that Thompson was wearing this patch will be discussed later in this opinion.
7. The FDA ultimately identified eight different kinds of defects, which ranged in severity from, e.g.,
"stringer leakers," where a strand of gel kept the seal on the patch from complete closure, to the defect at
issue herein, a "fold-over" defect, which left an entire side of the Duragesic patch unsealed.
8. For another court's discussion of the FDA's Establishment Inspection Report and its Inspectional
Observations, please see Miller v. Janssen Pharmaceutical Products, L.P., No. 05-CV-4076-DRH , 2007 U.S.
Dist. LEXIS 31863, at *8-9 (S.D. Ill. May 1, 2007) (not designated for publication).
9. The FDA has stated that: Recall is an effective method of removing or correcting consumer products that are in
violation of laws administered by the Food and Drug Administration. Recall is a voluntary
action that takes place because manufacturers and distributors carry out their responsibility
to protect the public health and well-being from products that present a risk of injury or gross
deception or are otherwise defective. 21 C.F.R.§ 7.40(a) (2009). A recall is an "alternative" to a FDA-initiated court action, and the agency, by
setting forth specific procedures, can "monitor recalls and assess the adequacy of a firm's efforts in recall."
Id. The FDA evaluates the health hazard presented by a product being recalled or considered for recall and
assigns the recall a classification, i.e., Class I, Class II, or Class III, to indicate the relative degree of health
hazard of the product being recalled or considered for recall: (1) Class I is a situation in which there is a reasonable probability that the use of, or exposure
to, a violative product will cause serious adverse health consequences or death. (2) Class II is a situation in which use of, or exposure to, a violative product may cause
temporary or medically reversible adverse health consequences or where the probability of
serious adverse health consequences is remote. (3) Class III is a situation in which use of, or exposure to, a violative product is not likely to
cause adverse health consequences. 21 C.F.R. § 7.41 (2009); 21 C.F.R.§ 7.3(m)(1)-(3) (2009). The recall of Alza's Duragesic patches was
originally classified as a Class I recall and ultimately reclassified as a Class II recall.
10. We note that some of the cases involving claims against Alza based on flaws in its Duragesic
fentanyl patches have also included claims against the prescribing doctor. See, e.g., Borowicz v. Alza Corp.,
No. 1:09-CV-00785, 2009 U.S. Dist. LEXIS 53589, at *1-3 (N.D. Ohio June 24, 2009); Browning v. Alza Corp.,
No. 1:09-04402009, 2009 U.S. Dist. LEXIS 42755, at *2 (S.D. W. Va. May 20, 2009). But see Middlebrooks
v. Johnson & Johnson Co., No. 4:08-CV-54 (CDL), 2008 U.S. Dist. LEXIS 75420, at *1-2 M.D. Ga. Aug. 26,
2008) (not designated for publication)(no claims raised against prescribing physician in context of suit against
Alza and other makers of allegedly defective Duragesic fentanyl patches); Workman v. Johnson & Johnson,,
No. 06-2523 (DMC), 2007 U.S. Dist. LEXIS 46214, at * 1-2 (D. N.J. June 26, 2007) (not designated for
publication) (same). In this case, neither party raised any allegation against the prescribing physician.
11. Less than a month passed from Dr. Benhamou's prescription of the Duragesic patches and
Thompson's death. During that same general time period, Alza logged 21 complaints about leaking Duragesic
patches from one particular lot in circulation within the health care supply system.
12. An electrocardiogram, or EKG, graphically records the electrical activity of the heart. See Columbia
Med. Ctr. of Las Colinas, Inc. v. Hogue, 271 S.W.3d 238, 244 (Tex. 2008) (citing Cecil Essentials of
Medicine 47, 55 (Thomas E. Andreoli, M.D. et al. eds., 6th ed. 2004)). 13. The expected range of fentanyl in Thompson's blood, based on the strength of her Duragesic patch
prescription, was between 1.0 to 2.0 ng/ml, but the actual level in her blood was 11.4 ng/ml.
14. Evidence at trial, including testimony from William Parks, the director of trade relations for Janssen,
established that Alza manufactures the Duragesic patch and Janssen sells and markets the Duragesic patch.
Alza and Janssen are both subsidiaries of Johnson & Johnson.
15. We note that the jury was instructed that evidence regarding adverse events was relevant on the
issue of notice to Alza, but not on causation or for any other purpose.
16. Testimony at trial indicated that the absorption of fentanyl from a damaged patch may be increased,
even without the patch system in place, where, for example, clothes cover the patch and act as an occlusive
layer, or the patient is laying on the patch, thus causing less evaporation of the vehicle medium and a greater
absorption of fentanyl. Dr. Benet further testified that temperature increases the rate of fentanyl absorption.
17. We note that the foreseeable risk of harm due to a misuse of the product, rather than an intended
use, is not an absolute bar to liability for that portion of an injury caused by a product's defective design;
instead, misuse of a product is a factor that must be considered in allocating responsibility for the injury.
Hernandez v. Tokai Corp., 2 S.W.3d 251, 258 (Tex. 1999).
18. Dr. Benhamou testified that the Duragesic prescriptions came with five patches in each box with
directions to change the patches every 72 hours. Dr. Benhamou testified that he prescribed a box of five
Duragesic 50 mcg patches to Thompson on January 13. On January 26, Dr. Benhamou increased the dosage
and prescribed a box of five Duragesic 50 mcg patches and a box of five 25 mcg patches. On February 11,
Thompson was given a prescription for a box of five 75 mcg patches. Utilizing a calendar and medical
records, Dr. Benhamou testified that the prescription dates indicated proper usage of the product and the
timing indicated usage of the 75 mcg patch at the time of Thompson's death.
19. While Thompson's daughter testified that she was wearing the patch on her stomach, and medical
records and testimony indicate that she was wearing the patch on her back, this discrepancy is not material
to the issue of whether or not Thompson properly used the product.
20. As explained at trial, the concept of postmortem drug redistribution is based on data suggesting that
there is a post-mortem diffusion of drugs along a concentration gradient, from sites of high concentration in
solid organs and central blood vessels, to sites of the lowest concentration in peripheral vessels such as the
subclavian and femoral veins. Postmortem redistribution of a drug may be the basis for elevated or toxic drug
concentrations after death, and accordingly, in certain cases, postmortem drug concentrations may not
accurately reflect antemortem drug levels.
21. Accu-Chem initially tested a sample of Thompson's blood for fentanyl in April; however, that test
failed. Dr. Tarver and Dr. Wimbish, both testified that this initial test was not valid and the results were not
reliable because the sample of blood provided for testing was insufficient and Accu-Chem did not have the
"amount of specimen that our protocol requires." Accu-Chem received six-tenths of a milliliter of blood,
whereas the standard operating procedure called for four milliliters, and the resulting analytical data did not
allow for forensic results. Dr. Jones, testifying on behalf of Alza, stated that Accu-Chem's first test failed
because the entire calibration failed "miserably" and the failure had "absolutely nothing" to do with the amount
of the sample. We need not further consider this issue because the results of this initial test are not at issue
in this case. We do note that Dr. Tarver stated that the testing methodology utilized by Accu-Chem did not
change between the two tests. 22. William Parks, the director of trade relations for Janssen and the liaison between retail and
wholesale customers and the company, was unable to determine which lot that the patch Thompson was
wearing came from.
23. See Torrington Co. v. Stutzman, 46 S.W.3d 829, 838-39 (Tex.2000) (discussing circumstances
when a defendant's post-sale defect investigation gives rise to a duty to perform that undertaking with ordinary
care). As we recognized years ago, Texas law is generally not the source of any duty to warn of product
defects uncovered after sale, but there is such a duty when the manufacturer has regained control of a
product after the defect was learned and the manufacturer failed to remedy a defect before reselling the
product. See Bell Helicopter Co. v. Bradshaw, 594 S.W.2d 519, 531-32 (Tex. Civ. App.-Corpus Christi 1979,
writ ref'd n.r.e.). Although Texas law is not the source of a more general post-sale duty, there are other
potential sources of such a duty which might be enforced under Texas law. See, e.g., Lowe v. Gen. Motors
Corp., 624 F.2d 1373, 1379-1380 (5th Cir. 1980) (violation of Federal Motor Vehicle Safety Standards can
serve as basis for negligence per se claim).
24. Likewise, Thompson's Duragesic patch was made according to the same control processes as
those defective patches which escaped detection and were shipped to the public with flawed seals despite
Alza's visual inspection of the patches as they passed along a conveyor belt.
25. As part of the Thompson family's burden on specific causation, they were required to offer evidence
excluding other causes of Michaelynn's death with reasonable certainty. Merrell Dow Pharms. v. Havner, 953
S.W.2d 706, 720 (Tex. 1997).
26. We note that we are not the first court to review Alza's claim that its own experts' dismissive
testimony about the flaws in its Duragesic fentanyl patches is authoritative or Alza's arguments that the
claimant's experts' contrary opinions should be wholly disregarded. See, e.g., Kunnemann v. Janssen
Pharmaceutica Products, L.P., No. 05 C 3211, 2008 U.S. Dist. LEXIS 101724, at *8-32 (N.D. Ill. Dec. 2, 2008)
(not designated for publication). Likewise, we are not the first court to conclude "[t]hese issues must be
resolved by a jury." See id. at *32.
27. In this context, the determination whether expert testimony is necessary is reviewed under a de
novo standard of review. See FFE Transp. Servs. v. Fulgham, 154 S.W.3d 84, 90 (Tex. 2004).
Palmer v. AH Robins Co., Inc. ( 1984 )
Daubert v. Merrell Dow Pharmaceuticals, Inc. ( 1993 )
Bic Pen Corp. v. Carter Ex Rel. Carter ( 2008 )
Norstrud v. Trinity Universal Insurance Co. ( 2003 )
Gulf States Utilities Co. v. Dryden ( 1987 )
J.K. & Susie L. Wadley Research Institute & Blood Bank v. ... ( 1992 )
Wright v. General Motors Corp. ( 1986 )
roy-ronald-lowe-as-administrator-of-the-estate-of-lou-ann-lowe-deceased ( 1980 )
Turner v. General Motors Corp. ( 1979 )
GTE Southwest, Inc. v. Bruce ( 1999 )
Caterpillar, Inc. v. Shears ( 1995 )
Turbines, Inc. v. Dardis ( 1999 )
Plas-Tex, Inc. v. U.S. Steel Corp. ( 1989 )
Whirlpool Corp. v. Camacho ( 2009 )
Shaun T. Mian Corp. v. Hewlett-Packard Co. ( 2007 )